Results to be expected
At room temperature, the measurement results for 95% of all control solution
tests should lie within the results range printed on the test strip container.
Warning
The results range printed on the container with the test strips applies only to
the control solution.
This is not a recommended value for your blood sugar
content.
If measurement results are outside the specified range, check the following pos-
sible causes:
Cause
Remedy
The first drop of control solution
was not disposed of.
•
The tip of the bottle was not wiped
clean.
•
The bottle was not shaken vigour-
ously enough.
Eliminate the cause and repeat the test.
The control solution has expired or is
contaminated.
Repeat the test with a
new bottle of control solution.
The control solution, test strip or mea-
suring device is too warm or too cold.
Bring the control solution and measuring
device to room temperature (+20°C to
+25°C) and repeat the test.
Test strip is damaged.
Repeat the test with a new test strip.
Test strip is out of date.
Open a new container of test strips.
Repeat the test.
There is a problem with the measuring
device.
Please contact customer service.