9 Troubleshooting
Messages on the display for batteries and blood sugar measurement
No. Cause
Action
E-b Flat batteries.
Replace all batteries.
E-t
Temperature of the measurement
room, measuring device or test strip
has exceeded the allowable limit.
Repeat test with new test strip as soon
as the measurement room, measuring
device and test strip have reached room
temperature (+20°C to +25°C).
E-U Used
or
contaminated
test strip was inserted.
•
Insert test strip which has not been
used or spoiled.
•
Repeat blood sugar measurement.
E-F Test strip was removed from the unit
during measurement.
Repeat blood sugar measurement.
Messages on the display for blood pressure measurement
No.
Cause
Action
Err 01 Cuff is not tight or the pressure
build-up is too low.
Put the cuff on correctly (see "Putting on
the inflatable cuff", page 25) and repeat
the measurement.
Err 02 Systolic pressure indeterminable. Repeat measurement without moving
and without speaking.
Err 03 Cuff pressure is too high (greater
than 300 mmHg).
Do not squeeze the cuff.
Err 04 Diastolic pressure indeterminable.
Repeat measurement without
moving and without speaking.
Err 05 Deflation rate is too high.
Verify that the cuff is correctly connected
to the unit and repeat the measurement.
Contact our customer service depart-
ment if the problem persists.