7 Main Menu Device Setup
LB 478 MPLM
68
56925-8BA2 Rev.00, 07/2019
Perform calibration
j
DANGER
Danger to life from electric shock!
u
The repair may only be carried out by a qualified electrician.
u
Please adhere to the relevant safety regulations.
u
During servicing on the hardware as well as during wiring of the detector,
the measuring system, connected relay contacts and all inputs and outputs
must be de-energised.
In case of an electric shock, carry out first aid measures and immediately call an emer-
gency service.
1.
Connect the measuring lines of the current measuring device on the back of
the EVU to the analog output.
2.
Observe the terminal allocation in the document “Technical information”.
3.
Click on the button
<Calibration>
.
u
The device switches to test mode and a new window (Analog output calibra-
tion) opens.
u
The calibration point 4 mA is displayed and the current measuring instrument
shows a value.
4.
Enter the indicated value on the current measuring instrument in the input
box (Fig. 49, item 4).
5.
Click on the button
<Continue>
.
u
The calibration point 20 mA is displayed and the current measuring instrument
shows a value.
6.
Enter the indicated value on the current measuring instrument in the input
box (Fig. 49, item 5).
7.
Click on the button
<Continue>
.
u
A message appears “Calibration successful”.
8.
Click on the button
<Continue>
.
u
The calibration of the analog output is completed.
Summary of Contents for Duo Xpert LB 478 MPLM
Page 2: ......
Page 4: ......
Page 12: ...1 About this Operating Manual LB 478 MPLM 12 56925 8BA2 Rev 00 07 2019...
Page 16: ...2 Safety LB 478 MPLM 16 56925 8BA2 Rev 00 07 2019...
Page 26: ...4 Installation LB 478 MPLM 26 56925 8BA2 Rev 00 07 2019...
Page 34: ...5 Electric Installation LB 478 MPLM 34 56925 8BA2 Rev 00 07 2019...
Page 40: ...6 Operation of the Software LB 478 MPLM 40 56925 8BA2 Rev 00 07 2019...
Page 74: ...7 Main Menu Device Setup LB 478 MPLM 74 56925 8BA2 Rev 00 07 2019...
Page 88: ...9 Troubleshooting LB 478 MPLM 88 56925 8BA2 Rev 00 07 2019...
Page 94: ...10 Maintenance and Repair LB 478 MPLM 94 56925 8BA2 Rev 00 07 2019...
Page 119: ...EmulsionSENS Technical Information Technische Information 56925TI1M Rev No 00 07 2019...