7 Main Menu Device Setup
LB 478 MPLM
66
56925-8BA2 Rev.00, 07/2019
AO Limits
Device Setup | Setup | Output | Analog Output (AO) | AO Limits
By clicking on the input fields (Fig. 48, item 1, item 2), the values [mA] for the lower
and upper current limit can be set. In addition to the default value of 3.8 mA, the
lower current limit can be set to 0 mA in order to switch the current output from
4 ... 20 mA to 0 ... 20 mA.
1
2
Input field Lower current limit [mA]
Input field Upper current limit [mA]
Fig. 48
Analogue output (AO Limit)
NOTICE
If the measurement is operated according to Namur, the standard current values
of 3.8 or 20.5 mA must be maintained.