LB 478 MPLM
7 Main Menu Device Setup
56925-8BA2 Rev.00, 07/2019
65
AO Failure Mode
Device Setup | Setup | Output | Analog Output (AO) | AO Failure Mode
The alarm function is set when an error is detected at the current output in the
window “Error mode”.
1
2
3
4
Selection arrow Alarm function
Live target value [mA]
Live actual value [mA]
Selection of the alarm function
Fig. 47
Analog output (AO Failure Mode)
The following behaviour of the current output can be assigned in case of error:
Namur High
22 mA (in error mode).
Namur Low
2 mA (in error mode).
Hold Value
Last value before the error.
User-defined value
The value can be set manually.
NOTICE
If the value "Hold Value" is set, it is recommended that the error relay is con-
nected in order to allow device errors to be transmitted to the control system.
Summary of Contents for Duo Xpert LB 478 MPLM
Page 2: ......
Page 4: ......
Page 12: ...1 About this Operating Manual LB 478 MPLM 12 56925 8BA2 Rev 00 07 2019...
Page 16: ...2 Safety LB 478 MPLM 16 56925 8BA2 Rev 00 07 2019...
Page 26: ...4 Installation LB 478 MPLM 26 56925 8BA2 Rev 00 07 2019...
Page 34: ...5 Electric Installation LB 478 MPLM 34 56925 8BA2 Rev 00 07 2019...
Page 40: ...6 Operation of the Software LB 478 MPLM 40 56925 8BA2 Rev 00 07 2019...
Page 74: ...7 Main Menu Device Setup LB 478 MPLM 74 56925 8BA2 Rev 00 07 2019...
Page 88: ...9 Troubleshooting LB 478 MPLM 88 56925 8BA2 Rev 00 07 2019...
Page 94: ...10 Maintenance and Repair LB 478 MPLM 94 56925 8BA2 Rev 00 07 2019...
Page 119: ...EmulsionSENS Technical Information Technische Information 56925TI1M Rev No 00 07 2019...