3 System Description
LB 478 MPLM
20
56925-8BA2 Rev.00, 07/2019
Status displays of the master EVU
The LEDs (Fig. 3, items 2-4) below the touch display show the current operating
status of the master EVU.
Display LED
Description
RUN This LED lights up green if the device is in operation
and fault-free. The current measurement is carried out.
RUN (flashing)
The RUN LED flashes green while the measurement is
(held) in the STOP state by user actions (e.g. stop func-
tion, simulation mode, plateau recording).
WARNING
WARNING This LED lights up yellow when a system
event of the type "Outside of specification", "Mainte-
nance required" or "Function check" is present.
All system events are described in chapter 9.2.
ERROR
ERROR
This LED lights up red if a system event of the type "Fail-
ure" is present. The current measurement is retained.
Check the Device Setup.
All system events are described in chapter 9.2.
RUN / WARNING / ERROR flashing
All three LEDS flash during the system test which is per-
formed as part of the start up process.
Summary of Contents for Duo Xpert LB 478 MPLM
Page 2: ......
Page 4: ......
Page 12: ...1 About this Operating Manual LB 478 MPLM 12 56925 8BA2 Rev 00 07 2019...
Page 16: ...2 Safety LB 478 MPLM 16 56925 8BA2 Rev 00 07 2019...
Page 26: ...4 Installation LB 478 MPLM 26 56925 8BA2 Rev 00 07 2019...
Page 34: ...5 Electric Installation LB 478 MPLM 34 56925 8BA2 Rev 00 07 2019...
Page 40: ...6 Operation of the Software LB 478 MPLM 40 56925 8BA2 Rev 00 07 2019...
Page 74: ...7 Main Menu Device Setup LB 478 MPLM 74 56925 8BA2 Rev 00 07 2019...
Page 88: ...9 Troubleshooting LB 478 MPLM 88 56925 8BA2 Rev 00 07 2019...
Page 94: ...10 Maintenance and Repair LB 478 MPLM 94 56925 8BA2 Rev 00 07 2019...
Page 119: ...EmulsionSENS Technical Information Technische Information 56925TI1M Rev No 00 07 2019...