LB 478 MPLM
7 Main Menu Device Setup
56925-8BA2 Rev.00, 07/2019
67
Calibrate
Device Setup | Setup | Output | Analog Output (AO) | AO Calibration
If there are any discrepancies between the target value and the actual value of the
current signal, then the current output may be calibrated again.
NOTICE
For calibration of the current output, an ammeter (not included in the scope of
delivery) is required, which is connected to the current output.
NOTICE
Berthold recommends calibrating the current outputs whenever a module has
been installed/replaced or if a software update has been carried out.
1
2
3
Execution with connection board
Execution with terminal block
Button
<Calibrate>
4
5
Input field 4 mA Calibration point
Input field 20 mA Calibration point
Fig. 49
Analog Output (Calibration)