7. Cleaning and Maintenance
LB 925 Crocodile miniWorkstation
Page 54 of 68
84024BA2 Rev. 03, 04/2020
7.5
Maintenance
7.5.1
Changing the Dispenser Lines
IMPORTANT
The user may exchange the dispenser lines if a tubing connection is
faulty or if desired, using the Dispenser Tubing Set coming with the
instrument. Refer to the Maintenance Schedule in chapter 7.1.
Operate as follows:
1.
Empty the dispenser lines.
2.
Turn the instrument off and disconnect it from the mains.
3.
Open the drawer at the side of the instrument. Unscrew the
screws with the provided Allen key only.
4.
Pull out the drawer carefully to make the tubing connections
at the dispenser pumps accessible.
5.
Unscrew the respective tubing’s fitting screw.
Figure 7-4:
Fitting screw at the dispenser pump
Figure 7-3:
Drawer at the side of the instrument
Fitting screw