LB 925 Crocodile miniWorkstation
4. Hardware Description
84024BA2 Rev. 03, 04/2020
Page 29 of 68
from time
to time. If a connection is leaking, it has to be replaced.
See chapter 7.5 for instructions.
4.4
Shaking Function
The Crocodile miniWorkstation features a shaking mode with
independent linear motion from 5 – 20 Hz. The function is controlled
by software.
4.5
Sample Incubation
The Crocodile miniWorkstation includes an incubation chamber and
provides a heating function from ambient +4°C up to 55°C. This
function is controlled via software.
When working with these functions, microplates with flat bottoms
are strongly recommended to achieve optimum heat transfer.
The microplate can be heated, but not cooled.
4.6
Washer
It is possible to connect up to three wash buffers to the inlets 1-3 at
the side of the instrument. The buffer lines are preinstalled. Use the
three square plastic bottles, coming with the instrument, for wash
buffers.
The washer manifold is built up of eight pairs of separate dispensing
needles (short needles) and aspiration needles (long needles). The
eight-way manifold can dispense in a range between 50 and 1000 µl.
Figure 4-6:
Position of washer inlets 1-3