7. Cleaning and Maintenance
84024BA2 Rev. 03, 04/2020
Page 53 of 68
the reagents in the lines for approximately 10 minutes. Empty
and dry the lines, then. Wipe the outside of all lines with 0.5%
bleach multiple times and with distilled water afterwards.
4.
Follow the description for cleaning the internal waste
container and the extraction hose but wipe out with 0.5%
bleach multiple times and with distilled water afterwards. Dry
the waste container and the extraction hose by pumping air.
5.
If necessary, decontaminate the reagent bottles and the waste
bottle with 0.5% bleach. Use distilled water afterwards and
dry them carefully.
You will find information on the Declaration on Decontamination in
chapter 11.3.1.