User Manual
DDT-063-000 Rev C Belmont Medical Technologies Page
87
of
106
The following messages should be checked and confirmed:
–
Low Core temperature thermoregulation is continuing…
–
Out of Normothermia Range
–
Patient Temperature above XX.X°C (*)
–
Patient Temperature below YY.Y°C (*)
–
Water Temp Too High (*)
NOTE:
Only authorized users can change the range of
the alarms marked by (*) in the Settings screen.
The user needs to insert a password to enter the
Settings panel and change the alarm limit.
Loss of Power Indicator
In the event that the machine loses power or is unplugged during
operation, a yellow indicator on the front panel will flash.
This indicator will continue to flash for 10 minutes until power is restored
or the Power Loss Alarm Disable button on the back of the unit is
pressed.
Alarm Delay
The following conditions generate an alarm only after a period of 30
seconds of being out of alarm limits:
–
Patient temperature is below XX.X°C:
Patient core temp is below the pre-configured alarm limit in
the Settings menu.
–
Patient temperature is above XX.X°C:
Patient core temp is above the pre-configured alarm limit in
the Settings menu.
–
Water temp too high:
Water temp is above the pre-configured alarm limit in the
Settings menu.
NOTE:
The limits are subject to change according to user settings.
Entering the limits range disables the alarms immediately. A new
alarm is generated again after an additional 30 seconds after the out
of the limit value is measured.