Table 12: Allon
®
System Messages Troubleshooting Guide (Con’t)
Message
Cause of Problem
Action to be
taken
Comments
Water temp. too High
The alarm for High
Water Temperature can
be configured in
"Settings". The alarm
and message are
activated according to
the selected alarm limit.
The available values
range from 36-42° in
increments of 0.5°C:
37°C, 38°C, 39°C,
40°C, 41°C, and 42°C.
Thermoregulation
stops until the water
cools or the system
halts.
Turn OFF the system
for 3 seconds and
then turn it ON
again. If the problem
persists,
turn OFF the Allon
®
and contact your
local service
representative.
The alarm can be muted
for 10 minutes.
Patient’s temperature is above
XX.X°C
The alarm for High
patient’s temperature
can be configured in
“Settings”.
The alarm and message
are activated according
to the selected alarm
limit.
The available values
range from 38-41° in
increments of 0.5°C:
38°C, 38.5°C, 39°C,
39.5°C, 40°C, 40.5°C,
and 41°C.
Check that the Core
Temperature Probe is
in place and follow
the patient’s
temperature.
Inform the physician.
Thermoregulation
continues.
The alarm can be muted
for 10 minutes.
Patient’s temperature is below
XX.X°C
The alarm for Low
Patient Temperature
can be configured in
“Settings”. The alarm
and message are
activated according to
the selected alarm
limit.
The available values
range from 30-35° in
increments of 0.5°C:
30°C, 30.5°C, 31°C,
31.5°C, 32°C, 32.5°C,
33°C, 33.5°C, 34 °C,
34.5°C, and
35 °C.
Check that the Core
temperature probe is
in place and follow
the patient’s
temperature.
Inform the physician.
Thermoregulation
continues.
The alarm can be muted
for 10 minutes.