Allon
®
User Manual
DDT-063-000 Rev C Belmont Medical Technologies
Page
66
of
106
Table 9: Inspection and Maintenance Schedule
Frequency
Inspection/Service
Performed By
Before each treatment
•
Clean connecting tubes and Quick
Coupling Connectors with a wet
cloth.
•
Perform a visual inspection for any
mechanical failure in sensors,
connecting tubes, and power
cable.
•
Perform a visual inspection of the
exterior of the Allon® Device.
•
Circulate NaDCC per
manufacturer’s instructions
through Allon® Device for 30
minutes.
Staff
As required by hospital/clinic
protocol
•
Routine external cleaning
and disinfecting.
•
Empty Allon:
–
For Frequent Use (3-4 times a
week): drain once a week
–
For Infrequent Use: drain
after each use
•
Replace Connecting Water
Hoses (PN #200-00109)
periodically
Staff
Annually
•
Thermal disinfection
•
Replace filter *
•
Preventive maintenance
Belmont Medical
Technologies’
authorized
technician
Cleaning and Disinfection
The cleaning and disinfecting of the Allon
®
includes both external and
internal cleaning and disinfection.
NOTE:
All instructions regarding the reusable temperature
probes are not relevant for US market or other select
markets.