Allon
®
User Manual
DDT-063-000 Rev C Belmont Medical Technologies Page
100
of
106
HH:MM format).
2. Press ENTER.
The cursor moves to the selected time spot and the Temperatures displayed
are the temperatures of the new spot.
2.
To move the cursor line, in time (X direction):
1. Click the Cursor
icon.
2. Bring the + to the cursor location, The + will convert to a double line .
3. Use the mouse to move the double line to a new cursor location.
NOTE:
The values of the temperature at the cursor
location appear in the window adjacent to the
curve color window.
3.
To bring cursor to the center of the graph:
1. Click
.
Modes and Error Area
This area provides the following information:
–
System mode
marked by letters (See Table 16) and a vertical line.
–
Rewarming
steps between 0º
C
and 0.5º
C
shown in the
example in pink (the step was first 0.4ºC and then changed to
0.2ºC).
–
Error
: Period with no control, in the example due to system
pause (yellow markings).
Table 16: Mode Codes
Code
Indicates
A
PowerUp
Cooling
Adult
B
PowerUp
Cooling
Neonate
C
PowerUp
Warming
Adult
D
PowerUp
Warming
Neonate
E
PowerUp
Rewarm
Adult
Figure 52: Example of “Modes and Error Area”