DDT-063-000 Rev C Belmont Medical Technologies
Page
69
of
106
4. Turn off the system and turn off the LED by pressing the Power Loss Alarm
Disable button.
5. Unplug the power cord from wall outlet.
6. Using a lint free cloth with sterile water, clean the exterior of the machine, the
LCD screen, the hoses, the power cord and the reusable temperature probes
1
from any soiling.
7. Prepare the disinfectant solution as described by the manufacturer.
8. Using a lint free cloth with the disinfectant, disinfect the exterior of the machine,
the LCD screen, the hoses, the power cord and the reusable temperature
probes
1
.
NOTE:
Allow disinfectant contact time according to the
disinfectant manufacturer's instructions.
9.
For residue removal, use a new lint free cloth moistened with sterile water.
Use the cloth on the exterior of the machine, the LCD screen, the power
cord, the hoses and the reusable
temperature probes
1
.
CAUTION!
Do not apply physical pressure on the screen.
10. Store the machine in a cool and dry place.
1.
Reusable probes - Not applicable for US market or other select markets