SPECIFICATIONS
This chapter lists and describes the technical specifications for the Allon
®
system and
CliniLogger™ accessory.
Allon
®
Technical Specifications
Allon
®
, one of Belmont Medical Technologies’ patient temperature management solutions, is a
servo-controlled, non-invasive thermal regulation system. Allon’s algorithm-driven heat pump
supplies heated water through the ThermoWrap® disposable patient wrap.
Control Unit
Physical Dimensions
Mobile Unit with 4 wheels and 2 brakes
260 mm W x 625 mm D x 940 mm H (10.23” W x 24.6” D x 37”H)
Net Weight
34 kg / 75 lb
Environmental Operating Conditions
Temperature
5°C to 40°C (41-104°F)
Humidity
10 to 93%, non-condensing
Note:
Do not use in an atmosphere with flammable anesthetic mixtures.
Environmental Storage Conditions
Temperature
-15°C to +68°C (5-154°F)
Humidity
10 to 93%, non-condensing
Hardware
Electricity Input Power
230/115 VAC (Switchable) with isolation transformer 50/60 Hz
Maximum Power
Consumption
690 Watts
230 VAC 3.0A
115 VAC 5.8A
Heat Exchangers
Peltier Technology - Thermoelectric Coolers (TECs)
External Ports
(1) Isolated Serial Port
LCD Display Size
144.8mm / 5.7” color display
LCD Display Resolution
320x240
User Interface
MultiCapacitive Touch Screen
5 soft push buttons
System Sensors
3 Internal Temperature Sensors:
1) Water In, 2) Water Out, and 3) Thermostat
2 Pressure Sensors
Water
Water Type:
Sterile or 0.22 µ filtered water
Tank Capacity:
6 liters (1.6 gallon)
Pump Rate:
1.2 L/minute
Water Temperature
Accuracy:
±0.3°C
Water Temperature
(Outflow) Range:
13-40.8°C (55.4-105.4°F)
Patient Temperature
Patient Temperature
Channels
2 channels:
1) Core and 2) Surface
Patient Temperature
Probe Accuracy
±0.3°C