background image

5.15  Total scale factor:  ‘5CALE-t’

Together  with  the  timebase  't-bA5E',  this  is  a
dividing  factor  that  defines  the  arithmetic
relationship  between  the  rate  and  total  displays.
When the flow rate and the total flow are required
in the same units 5CALE-t = 1.0  

The  total  scale  factor  may  be  set  to  any  value
between 0.0001 and 99999 allowing the total flow
and  rate  of  flow  to  be  displayed  in  different
engineering units. 

5CALE-t  =    Units of rate display

  

                   Units of total display

When  calculating  5CALE-t  the  required  units  of
measurement  for  the  total  display  must  be
converted  to  those  used  for  the  rate  display  as
shown in the following example:

The  BA558E  rate  display  represents  litres  per
second but the total flow display is required in UK
gallons.   There are 4.5461 litres in a UK gallon.

5CALE-t   =    Units of rate display
                        Units of total display

       =            1 litre 

                              1 gallon

=           1 litre        

  

           1 / 4.5461 litres 

5CALE-t   =    4.5461

Alternatively,  if  the  total  display  is  required  in
thousand of UK gallons

=           1 litre          

  

           1 / 4546.1 litres

5CALE-t   =    4546.1

The total scale factor 5CALE-t is not dependent on
the  rate  display  units  of  time.  i.e.  flow  units  per
second, minute or per hour.  The timebase t-bA5E
accounts for these differences.
To  check  or change the total  scale  factor select
'5CALE-t' from the configuration menu and press 

P

to  reveal  the  existing  setting,  one  digit  will  be
flashing.  The value of the flashing digit may be
changed by pressing the 

 or 

 button.  When

this digit is set as required pressing 

P

 will transfer

control to the next digit, which may be adjusted in
the same way.

To position the decimal point, repeatedly operate
the 

P

 button until the decimal point starts flashing.

The  decimal  point  can  then  be  positioned  by
pressing the 

 or 

 button.  

When the required total scale factor has been set,
press  

E

  t

o  enter  the  number  and  return  to  the

‘5CALE-t’ prompt in the configuration menu.

5.16  Clip-off:  ‘CLiP-oFF’

To prevent totalisation of very low flow rates, which
over  long  periods  may  result  in  significant
totalisation errors, the BA558E rate display can be
conditioned to stop totalisation when the flow rate
falls below a specified clip-off threshold.    Clip-off
is an absolute value, for bi-direction flow systems it
inhibits totalisation for flow rates bclip-off
and –clip-off.  

Clip-off  is  shown  in  the  same  units  as  the
instruments rate display. To check or change the
clip-off  threshold  select  'CLiP-oFF'  from  the
configuration  menu  and  press  

P

  to  reveal  the

current level, one digit will be flashing.  The value
of the flashing digit may be changed by pressing
the  

 or  

  button.     When   this   digit   is   set   as

required pressing 

P

 will transfer control to the next

digit,  which  may  be  adjusted  in  the  same  way.
When set as required, press the

 

E

 button to enter

the revised threshold and return to the ‘CliP-oFF’
prompt in the configuration menu.

For  uni-directional  flow  when  the  flow  rate  falls
below the clip-off threshold, totalisation is inhibited,
the  hold  annunciator  is  activated  but  the  flow
indicator continues to rotate until the input current
falls below 4mA.

For  bi-directional  flow  when  the  positive  or
negative flow rate is less than the absolute clip-off
threshold,  totalisation  is  inhibited  and  the  hold
annunciator  is  activated.    The  flow  indicator
continues to rotate until the low rate is less than
±0.05% of flow rate display span.

5.17 Local reset of total and grand total.
         ‘LoC. R5Et’

This function  contains  two sub-functions ‘t-rE5Et’
and ‘Gt-rE5Et’ which allow the total display and the
grand  total  to  be  independently  reset  using  the
instrument  push  buttons  without  access  to  the
configuration menu from the operating mode.

16

Summary of Contents for BA558E

Page 1: ...Issue 3 4th July 2014 BA558E General Purpose 4 20mA loop powered panel mounting Rate Totaliser Issue 3...

Page 2: ...ion using external current source 6 1 1 Example Adding break points to a new Rate Totaliser 6 2 Calibration using internal reference 6 2 1 Example Adding break points to a new Rate Totaliser 6 3 Error...

Page 3: ...vated Loading message displayed Configuration message displayed Model and firmware version displayed The BA558E Rate Totaliser starts functioning using configuration information stored in the instrume...

Page 4: ...mode Operating the or button will toggle the display to CLr YE5 with YE5 flashing operating the E button will then clear the grand total and Gt CLrd will be displayed to confirm the operation The ins...

Page 5: ...nce of less than 1k as shown in Fig 2 Note The negative reset terminal RS2 is internally connected to the negative 4 20mA input terminal 3 The total display may also be reset to zero by operating the...

Page 6: ...3 Installation Procedure a Cut the specified aperture in the panel To achieve an IP66 seal between the instrument enclosure and the instrument panel the aperture must have the tighter tolerances spec...

Page 7: ...as an accessory from BEKA associates Custom printed scale cards can also be supplied To change a scale card unclip the protruding end of the flexible strip by gently pushing it upwards and pulling it...

Page 8: ...lay at 20mA are adjustable although for flow application the display at 4mA is usually zero The position of the decimal point in the rate display is defined by the dP function To simplify calibration...

Page 9: ...onfiguration Access code CodE 0000 Function FunC 5td Linear Rate display at 4mA Zero 0 00 Rate display at 20mA 5PAn 100 00 Resolution rE5n 1 digit P button in display mode C P Local reset of total t r...

Page 10: ...s operated in the display mode Select PC of 4 20mA input 4 20 Current in mA Note PC is not available with lineariser See section 5 14 Display Summary of function 5CALE t Total scale factor Defines the...

Page 11: ...ntains two sub functions Select ConF Returns instrument to default configuration shown in section 5 2 LtAb Returns lineariser to defaults shown in section 6 5 Both instructions must be confirmed by en...

Page 12: ...12...

Page 13: ...arity with bi directional flow Use of the lineariser is described in section 6 of this instruction manual 5 5 Display resolution rE5n This function defines the resolution of the least significant disp...

Page 14: ...y and identify the lower display variable but on the larger upper display The position of the decimal point can be positioned in the same way by operating the and push buttons when set as required ent...

Page 15: ...s been activated see section 5 17 To zero the grand total from within the configuration menu select Clr Gtot and press P which will cause the instrument to display Clr no Press the or button until CLr...

Page 16: ...he or button When the required total scale factor has been set press E to enter the number and return to the 5CALE t prompt in the configuration menu 5 16 Clip off CLiP oFF To prevent totalisation of...

Page 17: ...g the P button will transfer control to the next digit When all the digits have been adjusted press E to return to the CodE prompt in the configuration menu The revised security code will be activated...

Page 18: ...Lineariser calibration using an external current source This method allows direct calibration of the lineariser with an external current source and is the preferred method when traceability is require...

Page 19: ...19...

Page 20: ...llowing break point 6 1 1 Example Adding break points to a new Rate Totaliser When adding break points to a new Rate Totaliser or to a lineariser following resetting to the factory defaults using the...

Page 21: ...e E button The corresponding rate display at each of the break points can now be defined using the diSP sub function Using the and buttons select the diSP sub function and press P which will reveal th...

Page 22: ...er the root extractor or the lineariser are required The standard linear relationship should therefore be selected With FunC displayed press P which will reveal the existing function Using the or butt...

Page 23: ...the totaliser scaling factor In this example the rate display is in litres per minute but the total display is required in cubic metres 5CALE t Units of rate display Units of total display There are 1...

Page 24: ...and total flow in cubic metres with a resolution of 0 1 cubic metres Total is to be shown on the upper large display and rate on the lower smaller display Totalisation is to occur for flow rates betw...

Page 25: ...tly the following table may help to identify the cause of the failure Symptom Cause Check No display and no voltage between terminals 1 3 or 12 3 if optional backlight is loop powered No power supply...

Page 26: ...New Rate Totalisers are supplied with the tag or application information specified when ordered thermally printed onto the rear panel adjacent to the terminals This tag information is not visible fro...

Page 27: ...sed and current will only flow in one direction Ron less than 5 0 7V Roff greater than 1M Fig 12 Equivalent circuit of each alarm output The solid state output of each alarm may be used to switch any...

Page 28: ...28...

Page 29: ...ivated when the selected display rate or total equals the setpoint See section 9 3 5 Hi Lo Alarm function Defines the alarm function as High or Low See section 9 3 6 no nC Normally open or normally cl...

Page 30: ...y pressing the or button followed by the E button to return to the alarm menu 9 3 7 Alarm output status no nC Configures the solid state alarm output to be open no or to be closed nC in the non alarm...

Page 31: ...ate or total display depending upon which has caused the alarm alternates between the numerical value and the alarm reference ALr1 or ALr2 when the alarm output is activated If both alarm outputs are...

Page 32: ...to the display mode Fig 15 Setpoint adjustment from the display mode To adjust an alarm setpoint select 5Px1 or 5Px2 and press P which will reveal the current setting Each digit of the setpoint may b...

Page 33: ...backlight 9 4 1 Loop powering the backlight The backlight is loop powered by connecting it in series with the Rate Totaliser s 4 20mA input as shown in Fig 17 which increases the maximum Rate Totalis...

Reviews: