To  add  a  break-point  using  the  

 or  

 button

select 'SEt' from the configuration menu and press

P

 

which will result in the

 

'Add' sub-function prompt

being displayed.  To enter the sub-function press 

P

which will reveal the current break-point and the
total  number  of  break-points  which  have already
been entered.    When adding a break-point the
insertion position can be selected using the 

 and

 push  buttons  followed  by  

P

  push  button  to

insert  the  additional  break-point.    In  previously
uncalibrated  linearisers  each  new  break-point
should  be  added  in  front  of  the  highest  existing
break-point, if this sequence is not followed a 'FaiL'
message  will  occur  when  the  breakpoints  are
calibrated.  See example in section 6.2.1.

The delete break-point, sub-function 'dEL' operates
in exactly the same way as the 'Add' sub-function
described above.   Once within the ‘dEL’ function
each time the 

P

 button is pressed a break-point is

removed.  When  deleting  a  break-point  from  a
calibrated  Rate  Totaliser,  the  break-point  to  be
deleted can be selected using the 

 and 

 push

buttons.  The minimum number of break-point is 2,
break-points '0

:

1' and '1

:

1'.  

When  the  required  number  of  linearising  break-
point has been entered, return to the linearisation
sub-menu by pressing  

E

.  

The Rate Totaliser will

display the 'Add' or 'dEL' prompt depending upon
the  last  sub-function  used.    The  Rate  Totaliser
input  current  and  corresponding  rate  display  at
each break-point can now be entered using the ‘in’
and ‘diSP’ sub-functions.

Using the 

 or 

 button select 'in' from the sub-

menu  and  press  

P

  which  will  reveal  the  starting

point for the first segment '0

:

n', where n is the total

number of break-point entered. Press  

P

 and use

the 

 and 

 buttons and the  

P

 button to move

between  digits,  to  enter  the  input  current  in
milliamps at which the first break-point is required,
usually 4.000mA.   When set as required, press 

E

to return to the ‘0

:

n’ prompt from which the next

break-point can be selected using the 

 and 

buttons.  When the required break-point has been
selected press 

P

 and enter the Rate Totaliser input

current at which this break-point is required using
the 

 and 

 buttons and the  

P

 button to move

between  digits,.  Repeat  this  procedure  until  the
Rate Totaliser input current at all the break-points
has been defined and then return to the ‘in’ sub-
function by pressing the 

E

 button. 

The  corresponding  rate  display  at  each  of  the
break-points can now be defined using the ‘diSP’
sub-function    Using the 

 and 

 buttons select

the  ‘diSP’  sub-function  and  press  

P

 which  will

reveal  the  starting  point  for  the  first  break-point
'0

:

n', where n is the total number of break-points

entered. Press  

P

 and use the 

 and 

 buttons

and the 

P

 button to move between digits, to enter

the  required  rate  display  at  the  first  break-point.
When set as required, press 

E

 to return to the ‘0

:

n’

prompt  from  which  the  next  break-point  can  be
selected  using  the  

 or  

 buttons.    When  the

required  break-point  has  been  selected  press  

P

and  set  the  required  rate  display  at  this  break-
point. 

Repeat  this  procedure  until  the  Rate  Totaliser
display  at  all  the  break-points  has  been  defined
and  then  return  to  the  ‘SEt’  function  in  the
configuration menu by pressing the 

E

 button twice.

Error message

If  during  calibration  the  Rate  Totaliser
displays  a  'FAiL'  error  message  the
current at which the selected break-point
is being set is not above the proceeding
break-point or is not below the following
break-point.   

6.2.1  Example, Adding break-points to a new
          Rate Totaliser

When adding break-points to a new Rate Totaliser,
or to a lineariser following resetting to the factory
defaults  using  the  'LtAb'  function  described  in
section 5.19, each additional break-point should be
added  before  the  highest  existing  breakpoint.
The  first  additional  break-point  should  be  added
before the default break-point '1

:

1' which will result

in a display of '1

:

2'.   If more new break-points are

required,  using  the  

 button  select  the  new

highest  break-point  '2

:

2'  and  add  the  second

additional  break-point  by  operating  the  

P

 push

button which will result in a display of '2

:

3'.  Repeat

the sequence until the required number of break-
points has been entered.

6.3 Under and over-range

The  lineariser  does  not  change  the  under  and
over-range  indication  described  in  section  5.20.
At input currents below that specified for the first
break-point '0

:

n', the Rate Totaliser will continue to

use the specified slope of the first segment.  

At input currents above that specified for the last
break-point 'n:n', the Rate Totaliser will continue to
use  the  slope  specified  for  the  last  lineariser
segment.   

6.4  Lineariser default configuration

When the lineariser is reset to the factory defaults
using the ‘LtAb’ function described in section 5.19,
the defaults conditions are: 

  Input        Rate
current      display

  

First break-point           '0

:

1'        4mA          0.00

  Second break-point      '1:1'      20mA      100.00

21

Summary of Contents for BA558E

Page 1: ...Issue 3 4th July 2014 BA558E General Purpose 4 20mA loop powered panel mounting Rate Totaliser Issue 3...

Page 2: ...ion using external current source 6 1 1 Example Adding break points to a new Rate Totaliser 6 2 Calibration using internal reference 6 2 1 Example Adding break points to a new Rate Totaliser 6 3 Error...

Page 3: ...vated Loading message displayed Configuration message displayed Model and firmware version displayed The BA558E Rate Totaliser starts functioning using configuration information stored in the instrume...

Page 4: ...mode Operating the or button will toggle the display to CLr YE5 with YE5 flashing operating the E button will then clear the grand total and Gt CLrd will be displayed to confirm the operation The ins...

Page 5: ...nce of less than 1k as shown in Fig 2 Note The negative reset terminal RS2 is internally connected to the negative 4 20mA input terminal 3 The total display may also be reset to zero by operating the...

Page 6: ...3 Installation Procedure a Cut the specified aperture in the panel To achieve an IP66 seal between the instrument enclosure and the instrument panel the aperture must have the tighter tolerances spec...

Page 7: ...as an accessory from BEKA associates Custom printed scale cards can also be supplied To change a scale card unclip the protruding end of the flexible strip by gently pushing it upwards and pulling it...

Page 8: ...lay at 20mA are adjustable although for flow application the display at 4mA is usually zero The position of the decimal point in the rate display is defined by the dP function To simplify calibration...

Page 9: ...onfiguration Access code CodE 0000 Function FunC 5td Linear Rate display at 4mA Zero 0 00 Rate display at 20mA 5PAn 100 00 Resolution rE5n 1 digit P button in display mode C P Local reset of total t r...

Page 10: ...s operated in the display mode Select PC of 4 20mA input 4 20 Current in mA Note PC is not available with lineariser See section 5 14 Display Summary of function 5CALE t Total scale factor Defines the...

Page 11: ...ntains two sub functions Select ConF Returns instrument to default configuration shown in section 5 2 LtAb Returns lineariser to defaults shown in section 6 5 Both instructions must be confirmed by en...

Page 12: ...12...

Page 13: ...arity with bi directional flow Use of the lineariser is described in section 6 of this instruction manual 5 5 Display resolution rE5n This function defines the resolution of the least significant disp...

Page 14: ...y and identify the lower display variable but on the larger upper display The position of the decimal point can be positioned in the same way by operating the and push buttons when set as required ent...

Page 15: ...s been activated see section 5 17 To zero the grand total from within the configuration menu select Clr Gtot and press P which will cause the instrument to display Clr no Press the or button until CLr...

Page 16: ...he or button When the required total scale factor has been set press E to enter the number and return to the 5CALE t prompt in the configuration menu 5 16 Clip off CLiP oFF To prevent totalisation of...

Page 17: ...g the P button will transfer control to the next digit When all the digits have been adjusted press E to return to the CodE prompt in the configuration menu The revised security code will be activated...

Page 18: ...Lineariser calibration using an external current source This method allows direct calibration of the lineariser with an external current source and is the preferred method when traceability is require...

Page 19: ...19...

Page 20: ...llowing break point 6 1 1 Example Adding break points to a new Rate Totaliser When adding break points to a new Rate Totaliser or to a lineariser following resetting to the factory defaults using the...

Page 21: ...e E button The corresponding rate display at each of the break points can now be defined using the diSP sub function Using the and buttons select the diSP sub function and press P which will reveal th...

Page 22: ...er the root extractor or the lineariser are required The standard linear relationship should therefore be selected With FunC displayed press P which will reveal the existing function Using the or butt...

Page 23: ...the totaliser scaling factor In this example the rate display is in litres per minute but the total display is required in cubic metres 5CALE t Units of rate display Units of total display There are 1...

Page 24: ...and total flow in cubic metres with a resolution of 0 1 cubic metres Total is to be shown on the upper large display and rate on the lower smaller display Totalisation is to occur for flow rates betw...

Page 25: ...tly the following table may help to identify the cause of the failure Symptom Cause Check No display and no voltage between terminals 1 3 or 12 3 if optional backlight is loop powered No power supply...

Page 26: ...New Rate Totalisers are supplied with the tag or application information specified when ordered thermally printed onto the rear panel adjacent to the terminals This tag information is not visible fro...

Page 27: ...sed and current will only flow in one direction Ron less than 5 0 7V Roff greater than 1M Fig 12 Equivalent circuit of each alarm output The solid state output of each alarm may be used to switch any...

Page 28: ...28...

Page 29: ...ivated when the selected display rate or total equals the setpoint See section 9 3 5 Hi Lo Alarm function Defines the alarm function as High or Low See section 9 3 6 no nC Normally open or normally cl...

Page 30: ...y pressing the or button followed by the E button to return to the alarm menu 9 3 7 Alarm output status no nC Configures the solid state alarm output to be open no or to be closed nC in the non alarm...

Page 31: ...ate or total display depending upon which has caused the alarm alternates between the numerical value and the alarm reference ALr1 or ALr2 when the alarm output is activated If both alarm outputs are...

Page 32: ...to the display mode Fig 15 Setpoint adjustment from the display mode To adjust an alarm setpoint select 5Px1 or 5Px2 and press P which will reveal the current setting Each digit of the setpoint may b...

Page 33: ...backlight 9 4 1 Loop powering the backlight The backlight is loop powered by connecting it in series with the Rate Totaliser s 4 20mA input as shown in Fig 17 which increases the maximum Rate Totalis...

Reviews: