background image

5.8  Lower display:   ‘di5P-2’

When  only  one  display  is  required  this  function
allows the lower display to be disabled.  To reveal
if the lower display is activated select 'di5P-2' from
the  configuration  menu  and  press

 

P

.      If  set  as

required,  press  

E

  to  return  to  the  configuration

menu, or press the 

 or 

 button to toggle the

function  between  ‘on’  and  ‘oFF’.    When  set  as
required  press

 

E

  to  enter  the  selection  and  to

return to the configuration menu.

5.9  Position of the decimal points:   ‘dP’

This  function  defines the  position  of  the  decimal
point in both the rate and total displays.   

Rate display

The rate display can have up to five digits plus
sign,  the  decimal  point  may  be  positioned
between any of the digits, or may be omitted.
The  BA558E uses  

floating  point arithmetic

 so

moving the position of the rate display decimal
point will affect totalisation.  

Total display

The total display can have up to eight digits or
seven digits with a negative sign, the decimal
point may be positioned with up to five digits
after it, or it may be omitted.   

Notes:

a. The   position   of   the   decimal   point   in   the

Grand  Total  will  be  the  same  as  that
selected for the Total Display.

 
b. If  the  total  display  decimal  point  can’t  be

moved to the left it is likely that the existing
display  will  be  over-ranged  by  the
adjustment.    The  total  display  should  be
reset to zero, which should allow the decimal
point to be positioned as required.  

c. If the decimal point is moved to the right, the

existing  total  display  will  be  truncated,  not
rounded,  the  lost  precision  can  not  be
recovered.

Positioning the decimal points

To adjust the position of either the rate or total
display  decimal  point  select  'dP'  from  the
configuration menu and press  

P

.   The upper

display defined as the rate or total display by
function ‘di5P-1’ (section 5.7) will be activated
and  identified.    The  decimal  point  can  be
positioned  by  operating  the  

 or  

 push

button, when set as required enter the setting
and  return  to  the  conditioning  menu  by
operating the 

E

 button.  Alternatively, to position

the  lower  display  decimal  point  press  the  

P

button which will display and identify the lower
display variable, but on the larger upper display.
The   position   of   the   decimal   point   can   be

positioned in the same way by operating the 

and  

 push  buttons,  when  set  as  required

enter the setting and return to the conditioning
menu by operating the 

E

 button.

5.10  Calibration of the rate display using an
         external current source:   ‘CAL’

This function enables the zero and span of the rate
display to be adjusted using an external calibrated
current  source.    When  used  with  an  accurate
traceable  current  source  this  is  the  preferred
method of calibration.

Zero 

 

rate display with 4mA input

Span

 

rate display with 20mA input

To calibrate the rate display select 'CAL' from the
configuration menu and press 

P

.

  The BA558E will

display 'ZEro' which is a request for a 4mA input
current.  Adjust the external  current calibrator to
4.000mA and again press  

P

 

which will reveal the

current zero rate display.   The flashing digit of the
display can be changed by pressing the 

 or 

buttons,  when  set  as  required  pressing  

P

 will

transfer  control  to  the  next  digit.    When  all  the
digits  have  been  adjusted,  press  

E

  to  enter  the

new zero and return to the 'ZEro' prompt .

Pressing the 

 button will cause the BA558E to

display '5PAn' which is a request for a 20mA input
current.  Adjust the external  current calibrator to
20.000mA and again press 

P

 

which will reveal the

existing span rate display.   The flashing digit of the
display can be changed by pressing the 

 or 

buttons,  when  set  as  required  pressing  

P

 will

transfer  control  to  the  next  digit.    When  all  the
digits have been adjusted press 

E

 t

o enter the new

rate span and return to the '5PAn' prompt.  Finally
press 

E

 

again to return to the configuration menu.

Notes:

a. The  Rate  Totaliser  input  current  must  be

adjusted to the required value before the zero
and span functions are entered by pressing the

 button.

b. Rate Totalisers may be calibrated at currents

other than 4 and 20mA, within the range 3.8 to
21.0mA providing the difference between the
two  currents  is  greater  than  4mA.    If  these
conditions  are  not  complied  with,  the
instrument  displays  ‘FaiL’  and  aborts  the
calibration.

 

c. If  the  zero  current  is  greater  than  the  span

current  the  instrument  will  be  reverse  acting
i.e. an increasing input current will cause the
display to decrease.

14

Summary of Contents for BA558E

Page 1: ...Issue 3 4th July 2014 BA558E General Purpose 4 20mA loop powered panel mounting Rate Totaliser Issue 3...

Page 2: ...ion using external current source 6 1 1 Example Adding break points to a new Rate Totaliser 6 2 Calibration using internal reference 6 2 1 Example Adding break points to a new Rate Totaliser 6 3 Error...

Page 3: ...vated Loading message displayed Configuration message displayed Model and firmware version displayed The BA558E Rate Totaliser starts functioning using configuration information stored in the instrume...

Page 4: ...mode Operating the or button will toggle the display to CLr YE5 with YE5 flashing operating the E button will then clear the grand total and Gt CLrd will be displayed to confirm the operation The ins...

Page 5: ...nce of less than 1k as shown in Fig 2 Note The negative reset terminal RS2 is internally connected to the negative 4 20mA input terminal 3 The total display may also be reset to zero by operating the...

Page 6: ...3 Installation Procedure a Cut the specified aperture in the panel To achieve an IP66 seal between the instrument enclosure and the instrument panel the aperture must have the tighter tolerances spec...

Page 7: ...as an accessory from BEKA associates Custom printed scale cards can also be supplied To change a scale card unclip the protruding end of the flexible strip by gently pushing it upwards and pulling it...

Page 8: ...lay at 20mA are adjustable although for flow application the display at 4mA is usually zero The position of the decimal point in the rate display is defined by the dP function To simplify calibration...

Page 9: ...onfiguration Access code CodE 0000 Function FunC 5td Linear Rate display at 4mA Zero 0 00 Rate display at 20mA 5PAn 100 00 Resolution rE5n 1 digit P button in display mode C P Local reset of total t r...

Page 10: ...s operated in the display mode Select PC of 4 20mA input 4 20 Current in mA Note PC is not available with lineariser See section 5 14 Display Summary of function 5CALE t Total scale factor Defines the...

Page 11: ...ntains two sub functions Select ConF Returns instrument to default configuration shown in section 5 2 LtAb Returns lineariser to defaults shown in section 6 5 Both instructions must be confirmed by en...

Page 12: ...12...

Page 13: ...arity with bi directional flow Use of the lineariser is described in section 6 of this instruction manual 5 5 Display resolution rE5n This function defines the resolution of the least significant disp...

Page 14: ...y and identify the lower display variable but on the larger upper display The position of the decimal point can be positioned in the same way by operating the and push buttons when set as required ent...

Page 15: ...s been activated see section 5 17 To zero the grand total from within the configuration menu select Clr Gtot and press P which will cause the instrument to display Clr no Press the or button until CLr...

Page 16: ...he or button When the required total scale factor has been set press E to enter the number and return to the 5CALE t prompt in the configuration menu 5 16 Clip off CLiP oFF To prevent totalisation of...

Page 17: ...g the P button will transfer control to the next digit When all the digits have been adjusted press E to return to the CodE prompt in the configuration menu The revised security code will be activated...

Page 18: ...Lineariser calibration using an external current source This method allows direct calibration of the lineariser with an external current source and is the preferred method when traceability is require...

Page 19: ...19...

Page 20: ...llowing break point 6 1 1 Example Adding break points to a new Rate Totaliser When adding break points to a new Rate Totaliser or to a lineariser following resetting to the factory defaults using the...

Page 21: ...e E button The corresponding rate display at each of the break points can now be defined using the diSP sub function Using the and buttons select the diSP sub function and press P which will reveal th...

Page 22: ...er the root extractor or the lineariser are required The standard linear relationship should therefore be selected With FunC displayed press P which will reveal the existing function Using the or butt...

Page 23: ...the totaliser scaling factor In this example the rate display is in litres per minute but the total display is required in cubic metres 5CALE t Units of rate display Units of total display There are 1...

Page 24: ...and total flow in cubic metres with a resolution of 0 1 cubic metres Total is to be shown on the upper large display and rate on the lower smaller display Totalisation is to occur for flow rates betw...

Page 25: ...tly the following table may help to identify the cause of the failure Symptom Cause Check No display and no voltage between terminals 1 3 or 12 3 if optional backlight is loop powered No power supply...

Page 26: ...New Rate Totalisers are supplied with the tag or application information specified when ordered thermally printed onto the rear panel adjacent to the terminals This tag information is not visible fro...

Page 27: ...sed and current will only flow in one direction Ron less than 5 0 7V Roff greater than 1M Fig 12 Equivalent circuit of each alarm output The solid state output of each alarm may be used to switch any...

Page 28: ...28...

Page 29: ...ivated when the selected display rate or total equals the setpoint See section 9 3 5 Hi Lo Alarm function Defines the alarm function as High or Low See section 9 3 6 no nC Normally open or normally cl...

Page 30: ...y pressing the or button followed by the E button to return to the alarm menu 9 3 7 Alarm output status no nC Configures the solid state alarm output to be open no or to be closed nC in the non alarm...

Page 31: ...ate or total display depending upon which has caused the alarm alternates between the numerical value and the alarm reference ALr1 or ALr2 when the alarm output is activated If both alarm outputs are...

Page 32: ...to the display mode Fig 15 Setpoint adjustment from the display mode To adjust an alarm setpoint select 5Px1 or 5Px2 and press P which will reveal the current setting Each digit of the setpoint may b...

Page 33: ...backlight 9 4 1 Loop powering the backlight The backlight is loop powered by connecting it in series with the Rate Totaliser s 4 20mA input as shown in Fig 17 which increases the maximum Rate Totalis...

Reviews: