5.2  Root extractor:   root

This function is primarily intended for use with
differential flowmeters which have a square law
4/20mA output.   To activate the square root
extractor select 'root' from the menu and press

  P

which will reveal if the function is 'On' or 'OFF'.   If
the function is set as required, press 

E

 to return to

the menu, or press the 

Up

 or 

Down

 button to

change the setting, followed by the

  E

 button to

return to the main menu.

For reference, the following table shows the output
current from a non-linearised differential flowmeter.

            

 % of full flow      Current output mA

 

2.5

4.01

10.0

4.16

25.0

5.00

50.0

8.00

75.0

13.00

100.0

20.00

5.3  Resolution:   rESn

This function defines the resolution of  the least
significant display digit.  Decreasing the display
resolution can improve the readability of a noisy
signal.   Select 'rESn' from the menu and press 

P

which will reveal the current display resolution.  To
change the resolution press the 

Up

 or 

Down

 button

to select 1, 2, 5 or 10 digits, followed by the 

E

button to return to the menu.  

5.4  Position of the decimal point:   d.P.

A dummy decimal point can be positioned between
any of  the digits or may be absent.  To position the
decimal point select 'd.P.' from the menu and press

.  The decimal point can then be moved or

turned off by pressing the 

Up

 or 

Down

 push-button,

followed by 

E

 to return to the menu.

5.5  Calibration of display using an external
       current source:   CAL

This function enables the zero and span of the
indicator to be adjusted using an external
calibrated current source.  When used with an
accurate traceable current source this is the
preferred method of calibration.

To calibrate the indicator select 'CAL' from the
main menu and press 

P.

  The indicator will display

'ZErO'  which is a request for a 4mA input current.
Adjust the external current calibrator to 4.000mA  
and again press 

 which will reveal the current

zero display.   Each digit of the indicator display
can be changed by pressing the 

Up

 or 

Down

buttons. When the first digit is correct,

pressing 

P

 will transfer control to the next digit.

When all the digits have been adjusted, press 

E

 to

enter the new zero.  The indicator will display 'Ent'
for a few seconds while the information is being
stored in memory, and will then return to the 'ZErO'
prompt .

To adjust the display at 20mA, press the 

Up

 button

which will cause the indicator to display 'SPAn'.
Adjust the external current calibrator to 20.000mA  
and again press 

 which will reveal the existing

full scale display.   Each digit of the indicator
display can be changed by pressing the 

Up

 or

Down

 buttons. When the first digit is correct,

pressing 

P

 will transfer control to the next digit.

When all the digits have been adjusted press 

E t

o

enter the new span.  The indicator will display 'Ent'
for a few seconds while the information is being
stored in memory, and will then return to the 'SPAn'
prompt.  Finally press 

E

 again to return to the main

menu.

This function may also be used when calibrating a
complete loop from primary element to indicator
display e.g. when the indicator is displaying the
output from a resistance thermometer transmitter.

5.6 Calibration using internal references:   SEt

This function enables the zero and span of the
indicator to be adjusted without the need for an
accurate external current source, or for
disconnection from the 4/20mA loop.  The
accuracy of this method depends upon the
accuracy of the internal references which should be
regularly calibrated as described in section 5.7.2 

To calibrate the indicator select 'SEt' from the main
 menu and press

 P.  

The indicator will display

'ZErO'; pressing 

P

 again will reveal the current

display at 4mA.   Each digit of the indicator display
can be changed by pressing the 

Up

 or 

Down

buttons. When the first digit is correct pressing 

P

will transfer control to the next digit.  When the
least significant digit has been adjusted, press 

E

 to

return to the 'ZErO' prompt which completes the
adjustment.

To adjust the display at 20mA, press the 

Up

 button

which will cause the indicator to display 'SPAn'.
pressing 

again will reveal the indicator display.   

Each digit can be changed by pressing the 

Up

 or

Down

 buttons. When the first digit is correct,

pressing 

P

 will transfer control to the next digit.

When the least significant digit has been adjusted
press 

E t

o return to the 'SPAn' prompt followed by

E

 to return to the menu.

8

Summary of Contents for BA524D

Page 1: ...BA524D loop powered 4 digit field mounting indicator Issue 4 3rd December 2003...

Page 2: ...7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Customer comments 8 Accessories 8 1 Units of measurem...

Page 3: ...l converter Fig 1 Simplified block diagram of BA524D Each time a 4 20mA current is applied to the instrument initialisation is performed After a short delay the following display sequence occurs 1 8 8...

Page 4: ...tput is greater than the sum of voltage drops introduced by the BA524D and the cable resistance Fig 3 Remote indication 4 INSTALLATION 4 1 Location The BA524D indicator is housed in a robust IP66 glas...

Page 5: ...ll the required glands conduit fittings or blanking plugs Note the temporary dust seals do not provide IP66 protection d Connect the field wiring to the terminals as shown in Fig 5 e Replace the instr...

Page 6: ...parameter root to be displayed If an incor rect code is entered the indicator will return to the display mode Once within the menu the required parameter can be reached by scrolling through the main...

Page 7: ...ction 50 or 60Hz may be selected This function must be set before the instrument is calibrated as it affects the indicator display See section 5 7 1 rEF Calibration of internal references The instrume...

Page 8: ...have been adjusted press E to enter the new zero The indicator will display Ent for a few seconds while the information is being stored in memory and will then return to the ZErO prompt To adjust the...

Page 9: ...tor to 20 000mA and again press P The indicator will display Ent when the 20mA reference has been updated and will then return to the 020A prompt Two operations of the E button will return the indicat...

Page 10: ...10...

Page 11: ...display the input current as a percentage of span when the P push button is operated Step 1 Connect indicator to calibrator Connect the indicator to an accurate external current source Terminal 1 posi...

Page 12: ...erences As in 7 1 the BA524D is required to display 50 0 with a 4mA input 1050 0 with a 20mA input from a linear transducer Maximum display resolution is required the frequency of the mains supply is...

Page 13: ...ng completion of calibration return to the display mode by pressing E 7 MAINTENANCE 7 1 Fault finding during commissioning If a BA524D fails to function during commissioning the following procedure sh...

Page 14: ...s also unscrew the two C screws securing the buttons and un plug the five way connector Finally unscrew all four D screws identified in Fig 4 and carefully lift off the front of the instrument The ins...

Page 15: ...hanumeric characters 10mm high or 1 row of 11 alphanumeric characters 7mm high or 2 rows of 18 alphanumeric characters 5mm high 8 2 Alarms The BA524D can be supplied with two solid state single pole a...

Page 16: ...Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 8 2 3 SP1 Alarm setpoint 1 Adjusts the alarm setpoint The alarm is activated when the indicator d...

Page 17: ...arm may be positioned anywhere between 19999 and 19999 providing this corresponds to an input current between 3 8 and 20 2mA e g If the indicator has been calibrated to display 0 with 4mA input and 10...

Page 18: ...et at 9000 and hysteresis of 200 will perform as follows High alarm will be activated when display equals or exceeds 9000 but will not reset until the display falls below 8800 8 2 8 Alarm delay dELA T...

Page 19: ...urity code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with this function disabled and the security code set to 000...

Page 20: ...lumetric units by filling the tank with known incremental volumes and calibrating the indicator to display the sum of the increments at each break point The number of break points required should firs...

Page 21: ...21...

Page 22: ...oint occurs Enables the required current at each break point to be defined without having to input an accurate input current to the indicator diSP Defines display at break point Enables the indicator...

Page 23: ...ayed 8 5 Display backlight The BA524D can be supplied with LED backlighting to improve display contrast when the indicator is installed in poorly illuminated areas The backlight is electrically isolat...

Reviews: