1.  DESCRIPTION

The BA524D is a 4½ digit loop powered digital
indicator which displays the current flowing in a
4/20mA loop in engineering units.  The indicator
only introduces a 1.1V drop which allows it to be
installed into almost any 4/20mA current loop.  No
additional power supply or battery is required.  

The BA524D is a third generation instrument
which, although having additional features
including a separate terminal compartment,
remains electrically compatible with the original
BA524 and BA524C.

The main application of the BA524D is to display a
measured variable or control signal in a process
area.  The zero and span of the display are
independently adjustable so that the indicator can
be calibrated to display any variable represented
by the 4/20mA current, e.g. temperature, flow,
pressure or level.

The instrument is housed in a robust IP66 glass
reinforced polyester (GRP) enclosure with an
armoured glass window.

2.  OPERATION

Fig 1 shows a simplified block diagram of a
BA524D.  The 4/20mA input current flows through
resistor R1 and forward biased diode D1.  The
voltage developed across D1, which is relatively
constant, is multiplied by a switch mode power
supply and used to power the instrument.  The
voltage developed across R1, which is proportional
to the 4/20mA input current, provides the input
signal for the analogue to digital converter.   

Fig 1  Simplified block diagram of BA524D

Each time a 4/20mA current is applied to the
instrument, initialisation is performed.  After a short
delay the following display sequence occurs:

-1.8.8.8.8

Display test in which all
segments of the display
are activated for 0.5
seconds.

Blank display

For 0.5 seconds.

Decimal points For 3 seconds.
cycled

Input current

Using calibration

display in

information stored in

engineering instrument 

memory.

units.

2.1  Controls

The BA524D is controlled and calibrated via four
push-button switches which are located behind the
instrument control cover, or as an option on the
instrument cover.    In the display mode i.e. when
displaying a variable, these switches have the
following functions:

P

While this button is pushed the indicator
will display the input current in mA, or
as a percentage of the instrument span
depending upon how the indicator has
been programmed.  When the button is
released the normal display in
engineering units will return.  The
function of this push-button is modified
when alarms or a tare function are fitted
to the indicator.

Down

While this button is pushed the indicator
will display the number which the
indicator has been calibrated to display
with a 4mA input.  When released the
normal display in engineering units will
return.

Up

While this button is pushed the indicator
will display the number which the
indicator has been calibrated to display
with a 20mA input.  When released the
normal display in engineering units will
return.

E

No function in the display mode.

3

Summary of Contents for BA524D

Page 1: ...BA524D loop powered 4 digit field mounting indicator Issue 4 3rd December 2003...

Page 2: ...7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Customer comments 8 Accessories 8 1 Units of measurem...

Page 3: ...l converter Fig 1 Simplified block diagram of BA524D Each time a 4 20mA current is applied to the instrument initialisation is performed After a short delay the following display sequence occurs 1 8 8...

Page 4: ...tput is greater than the sum of voltage drops introduced by the BA524D and the cable resistance Fig 3 Remote indication 4 INSTALLATION 4 1 Location The BA524D indicator is housed in a robust IP66 glas...

Page 5: ...ll the required glands conduit fittings or blanking plugs Note the temporary dust seals do not provide IP66 protection d Connect the field wiring to the terminals as shown in Fig 5 e Replace the instr...

Page 6: ...parameter root to be displayed If an incor rect code is entered the indicator will return to the display mode Once within the menu the required parameter can be reached by scrolling through the main...

Page 7: ...ction 50 or 60Hz may be selected This function must be set before the instrument is calibrated as it affects the indicator display See section 5 7 1 rEF Calibration of internal references The instrume...

Page 8: ...have been adjusted press E to enter the new zero The indicator will display Ent for a few seconds while the information is being stored in memory and will then return to the ZErO prompt To adjust the...

Page 9: ...tor to 20 000mA and again press P The indicator will display Ent when the 20mA reference has been updated and will then return to the 020A prompt Two operations of the E button will return the indicat...

Page 10: ...10...

Page 11: ...display the input current as a percentage of span when the P push button is operated Step 1 Connect indicator to calibrator Connect the indicator to an accurate external current source Terminal 1 posi...

Page 12: ...erences As in 7 1 the BA524D is required to display 50 0 with a 4mA input 1050 0 with a 20mA input from a linear transducer Maximum display resolution is required the frequency of the mains supply is...

Page 13: ...ng completion of calibration return to the display mode by pressing E 7 MAINTENANCE 7 1 Fault finding during commissioning If a BA524D fails to function during commissioning the following procedure sh...

Page 14: ...s also unscrew the two C screws securing the buttons and un plug the five way connector Finally unscrew all four D screws identified in Fig 4 and carefully lift off the front of the instrument The ins...

Page 15: ...hanumeric characters 10mm high or 1 row of 11 alphanumeric characters 7mm high or 2 rows of 18 alphanumeric characters 5mm high 8 2 Alarms The BA524D can be supplied with two solid state single pole a...

Page 16: ...Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 8 2 3 SP1 Alarm setpoint 1 Adjusts the alarm setpoint The alarm is activated when the indicator d...

Page 17: ...arm may be positioned anywhere between 19999 and 19999 providing this corresponds to an input current between 3 8 and 20 2mA e g If the indicator has been calibrated to display 0 with 4mA input and 10...

Page 18: ...et at 9000 and hysteresis of 200 will perform as follows High alarm will be activated when display equals or exceeds 9000 but will not reset until the display falls below 8800 8 2 8 Alarm delay dELA T...

Page 19: ...urity code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with this function disabled and the security code set to 000...

Page 20: ...lumetric units by filling the tank with known incremental volumes and calibrating the indicator to display the sum of the increments at each break point The number of break points required should firs...

Page 21: ...21...

Page 22: ...oint occurs Enables the required current at each break point to be defined without having to input an accurate input current to the indicator diSP Defines display at break point Enables the indicator...

Page 23: ...ayed 8 5 Display backlight The BA524D can be supplied with LED backlighting to improve display contrast when the indicator is installed in poorly illuminated areas The backlight is electrically isolat...

Reviews: