BEKA BA524D Manual Download Page 13

Step 5

Position dummy decimal point

Select 'd.P' from the main menu and
then press

 P

.  Using the 

Up

 and 

Down

push-buttons position the dummy
decimal point in front of the least
significant digit.
Press 

E

 to return to the main menu.

Step 6

Calibrate display

With any input current between 4 and
20mA select 'SEt' from the main menu
and press 

P.

   The indicator will display

'ZErO' in the sub-menu; press 

P

 to

reveal the existing zero display. Using
the 

Up

Down

 and 

P

 push-buttons enter

the required zero display of   -50.0
Press 

E

 to return to the 'ZErO' prompt.

Again with any input current between 4
and 20mA press the

 Up 

push-button

and the indicator will display 'SPAn'.
Press 

to reveal the existing span

display.  Using the 

Up

Down

 and 

P

push-buttons enter the required span
display of  1050.0  Press 

E

 to return to

the 'SPAn' prompt. Press 

E

 again to

return to the main menu.

Step 7

Return to display mode

Following completion of calibration
return to the display mode by pressing

'E

'.

7.  MAINTENANCE

7.1 Fault finding during commissioning

If a BA524D fails to function during commissioning
the following procedure should be followed:

Symptom          Cause                   Solution

No display

Incorrect

There should be 1V

wiring

between terminals 1 &
3 with terminal 1
positive.  

No display

Incorrect 

Check that a current

and no volts

wiring or

is flowing in the loop.

between

no power

terminals

supply.

1 and 3.

Insufficient

Check supply voltage 

loop voltage  and voltage drops
to operate 

caused by all

indicator

components in the loop.

Indicator 

Positive

The indicator has been 

displays 1

over-range

incorrectly calibrated &
is trying to display a 
number greater than
19999.

Indicator

Negative 

The indicator has been

displays -1

Over-range

incorrectly calibrated &
is trying to display a 
number less than
-19999.

Unstable

4/20mA input Check loop supply

display

has a large

voltage. 

ripple content.

Unable to  

Incorrect

Enter correct security

enter the

security 

code or fit security

programme

code 

link in override position.

mode

entered.

See Fig 7.

13

Summary of Contents for BA524D

Page 1: ...BA524D loop powered 4 digit field mounting indicator Issue 4 3rd December 2003...

Page 2: ...7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Customer comments 8 Accessories 8 1 Units of measurem...

Page 3: ...l converter Fig 1 Simplified block diagram of BA524D Each time a 4 20mA current is applied to the instrument initialisation is performed After a short delay the following display sequence occurs 1 8 8...

Page 4: ...tput is greater than the sum of voltage drops introduced by the BA524D and the cable resistance Fig 3 Remote indication 4 INSTALLATION 4 1 Location The BA524D indicator is housed in a robust IP66 glas...

Page 5: ...ll the required glands conduit fittings or blanking plugs Note the temporary dust seals do not provide IP66 protection d Connect the field wiring to the terminals as shown in Fig 5 e Replace the instr...

Page 6: ...parameter root to be displayed If an incor rect code is entered the indicator will return to the display mode Once within the menu the required parameter can be reached by scrolling through the main...

Page 7: ...ction 50 or 60Hz may be selected This function must be set before the instrument is calibrated as it affects the indicator display See section 5 7 1 rEF Calibration of internal references The instrume...

Page 8: ...have been adjusted press E to enter the new zero The indicator will display Ent for a few seconds while the information is being stored in memory and will then return to the ZErO prompt To adjust the...

Page 9: ...tor to 20 000mA and again press P The indicator will display Ent when the 20mA reference has been updated and will then return to the 020A prompt Two operations of the E button will return the indicat...

Page 10: ...10...

Page 11: ...display the input current as a percentage of span when the P push button is operated Step 1 Connect indicator to calibrator Connect the indicator to an accurate external current source Terminal 1 posi...

Page 12: ...erences As in 7 1 the BA524D is required to display 50 0 with a 4mA input 1050 0 with a 20mA input from a linear transducer Maximum display resolution is required the frequency of the mains supply is...

Page 13: ...ng completion of calibration return to the display mode by pressing E 7 MAINTENANCE 7 1 Fault finding during commissioning If a BA524D fails to function during commissioning the following procedure sh...

Page 14: ...s also unscrew the two C screws securing the buttons and un plug the five way connector Finally unscrew all four D screws identified in Fig 4 and carefully lift off the front of the instrument The ins...

Page 15: ...hanumeric characters 10mm high or 1 row of 11 alphanumeric characters 7mm high or 2 rows of 18 alphanumeric characters 5mm high 8 2 Alarms The BA524D can be supplied with two solid state single pole a...

Page 16: ...Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 8 2 3 SP1 Alarm setpoint 1 Adjusts the alarm setpoint The alarm is activated when the indicator d...

Page 17: ...arm may be positioned anywhere between 19999 and 19999 providing this corresponds to an input current between 3 8 and 20 2mA e g If the indicator has been calibrated to display 0 with 4mA input and 10...

Page 18: ...et at 9000 and hysteresis of 200 will perform as follows High alarm will be activated when display equals or exceeds 9000 but will not reset until the display falls below 8800 8 2 8 Alarm delay dELA T...

Page 19: ...urity code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with this function disabled and the security code set to 000...

Page 20: ...lumetric units by filling the tank with known incremental volumes and calibrating the indicator to display the sum of the increments at each break point The number of break points required should firs...

Page 21: ...21...

Page 22: ...oint occurs Enables the required current at each break point to be defined without having to input an accurate input current to the indicator diSP Defines display at break point Enables the indicator...

Page 23: ...ayed 8 5 Display backlight The BA524D can be supplied with LED backlighting to improve display contrast when the indicator is installed in poorly illuminated areas The backlight is electrically isolat...

Reviews: