8.  MAINTENANCE

8.1  Fault finding during commissioning

If  a  BA314NG  Tachometer  fails  to  function  during
commissioning  the  following  procedure  should  be
followed:

Symptom

Cause

Check:

No display

No power supply,

or incorrect wiring.

Note: Terminals 2,

6 & RS2 are

interconnected

within the

instrument. 

That there is

between 10 and

30V on  terminals 1

& 2 with terminal 1

positive.

Tachometer is

receiving power but

pulse input

indicator not

rotating

No input pulses,

incorrect input

configuration,

incorrect linking of

terminals 3 & 4. 

Input configuration.

Linking of

terminals 3 & 4.

That input signal

polarity is correct.

Pulse input

indicator rotating,

but incorrect speed

display

Incorrect speed

display calibration.

5CALE

 . 

5

t-bA5E

Pulse input

indicator rotating

but missing or

incorrect

run-time display.

Run-time display is

not activated.

Tachometer speed

display is less than

clip-off value.

Remote reset

switch contacts

closed.

di5P-2

 is activated

See 6.8

If HOLD

annunciator  is

activated, enter

smaller 

Clip-off

value. See 6.13

That ‘RESET’

annunciator is not

activated.  If it is,

check reset wiring

and switch.

Unstable

Tachometer

display.

Noisy pulse input

signal.

Eliminate source of

electrical noise.

Increase debounce

and/or display filter.

See 6.12

Unable to enter

configuration

menu.

Incorrect security

code.

That the correct

security code is

being used.

See 6.18

Contact BEKA if

code is lost.

Clip-off does not

function.

Clip-off has

automatically been

reset to zero

following

calibration change.

Reconfigure

 clip-off.

See 6.13

Alarms do not

function.

Alarms have been

disabled following

calibration change.

Re-enable both

alarms.

See 9.3.4

8.2  Fault finding after commissioning

EN 60079-17  

Electrical installations inspection and

maintenance

 permits live maintenance in Zone 2 if a

risk  analysis  demonstrates  that  this  does  not
introduce  an  unacceptable  risk.    The  removal  of
covers  [opening  of  Ex n  instrument  enclosure]  is
permitted if this can be done without contaminating
the interior of the instrument with dust or moisture.
Some  end-users  may  prefer  not  to  permit  live
maintenance to minimise risk.

ENSURE PLANT SAFETY BEFORE

STARTING MAINTENANCE

Live maintenance within the hazardous area
should  only  be  performed  when  it  is
permitted by risk analysis, or when there is
no  risk  of  a  flammable  atmosphere  being
present.

If  a  BA314NG  Tachometer  fails  after  it  has  been
functioning correctly, the following table may help to
identify the cause of the failure.

Symptom

Cause

Check:

No display

No power supply.

That there is

between 10 and

30V on terminals

1 & 2.

Pulse input

indicator not

rotating.

No input pulses.

Output from

sensor.

Wiring between

sensor and

Tachometer.

Pulse input

indicator rotating,

run-time display

not functioning.

HOLD annunciator

activated.

Input below 

clip-off threshold.

Adjust 

Clip-off

threshold.

Unstable rate

display

Noisy pulse input

signal

Locate source of

electrical noise, or

increase

debounce and

rate display filter.

If this procedure does not reveal the cause of the
fault,  it  is  recommended  that  the  instrument  is
replaced. 

8.3  Servicing

We recommend that faulty BA314NG Tachometers
are  returned  to  BEKA associates  or  to  your  local
BEKA agent for repair.

8.4  Routine maintenance

The  mechanical  and  electrical  condition  of  the
instrument should be regularly checked.  Inspection
frequency  should  be  adjusted  to  suit  the
environmental conditions.

28

Summary of Contents for BA314NG

Page 1: ...Issue 5 9th July 2019 BA314NG Ex nA and Ex tc Tachometer Issue 5...

Page 2: ...dAtE 6 8 Run time display di5P 2 6 9 Position of the decimal points dP 6 10 Speed scale factor 5CALE 5 6 11 Timebase t bA5E 6 12 Display filter FiLtEr 6 13 Clip off CLP oFF 6 14 Local reset LoC clr 6...

Page 3: ...larm silence time 5IL 9 3 12 Flash display when alarm occurs FL5H 9 3 13 Access Setpoint AC5P 9 3 14 Adjusting alarm setpoints from the display mode 9 4 4 20mA output 9 4 1 Ex nA certification 9 4 2 S...

Page 4: ...m compliance with the European ATEX Directive 2014 34 EU For international applications the BA314NG also has IECEx certification which is described in Appendix 2 For applications in the USA and Canada...

Page 5: ...the Tachometer may be configured to provide direct access to the alarm setpoints from the display mode when the and buttons are operated simultaneously See 9 3 13 and 9 3 14 2 3 Displays The BA314NG...

Page 6: ...the Type Examination Certificate This allows the Tachometer to be adjusted and configured live when installed in Zones 2 When connected to a suitable system the BA314NG Tachometer may be installed in...

Page 7: ...ety parameters Input Output Type of input Link 3 4 Ui Uo Io Switch contact Yes 15V 10 5V 9 2mA Proximity detector Yes 15V 10 5V 9 2mA Open collector Yes 15V 10 5V 9 2mA Magnetic pick off No 30V 1 1V 0...

Page 8: ...that terminals 2 6 and RS2 are interconnected within the BA314NG See Fig 1 4 1 Power supply The BA314NG Tachometer requires a minimum of 10V between terminal 1 2 and consumes 10mA without optional bac...

Page 9: ...eshold and rise above the upper thresholds shown in the following table Input transducer Switching thresholds Lower Upper Open collector 2k 10k Voltage pulse low 1 0V 3 0V Voltage pulse high 3 0V 10 0...

Page 10: ...d associated wiring should be able to withstand a 500V rms insulation test to earth The BA314NG contains a configurable debounce circuit to prevent false triggering See section 6 6 4 2 6 Voltage pulse...

Page 11: ...the bottom right hand corner the back box for interconnecting cable screens The BA314NG Tachometer may be pipe mounted using a BA393G pipe mounting kit 5 2 Installation Procedure Fig 6 illustrates the...

Page 12: ...instrument is ordered a blank scale card will be fitted which can easily be marked on site with a dry transfer or a permanent marker Custom printed scale cards are available from BEKA associates as a...

Page 13: ...factor that determines whether the Tachometer displays speed per second per minute or per hour The Tachometer incorporates a run time counter that displays the time in hours that the speed of the mon...

Page 14: ...on or oFF See section 6 8 Display Summary of function dP Decimal points Defines the position of the decimal point in the Tachometer speed display See section 6 9 5CALE 5 Speed scale factor 5CALE 5 is...

Page 15: ...6 Display Summary of function CLr Gtot Resets grand total run time to zero This function resets the grand total run time to zero from within the configuration menu when CLr YE5 is selected and 5urE is...

Page 16: ...16...

Page 17: ...17...

Page 18: ...ges such as those resulting from a mechanical contact closing and bouncing Three levels of protection may be selected and the amount of debounce applied depends upon the type of Tachometer input that...

Page 19: ...g shaft generates 18 pulses per revolution to produce a Tachometer speed display in revolutions 5CALE 5 should be set to 18 0 The units of the Tachometer speed display are pulses per unit of time The...

Page 20: ...sed it is not necessary to enter a clip off value see 6 8 To check or change the clip off threshold select CLP oFF from the configuration menu and press which will reveal the current setting The thres...

Page 21: ...isplayed and then press which will result in a 0000 prompt being displayed with the first digit flashing This is a request for the instruction to be confirmed by entering 5urE using the or button to a...

Page 22: ...the wires entering the hazardous area should be individually fused and contain a means of isolation These two requirements may be satisfied by installing DIN rail mounted terminals incorporating a fus...

Page 23: ...ting on or oFF The function can be changed by pressing the or push button followed by the button to return to EnbL prompt in the sub menu 6 20 6 Source of output pulse 5ourCE The output pulse may be d...

Page 24: ...to select the required value followed by the button to return to durAtion prompt in the sub menu Note This function only appears in the pulse output sub menu when 5CALEd is selected in the 5ource sub...

Page 25: ...al between display updates Using the or button select uPdAtE in the configuration menu and press to reveal how frequently the Tachometer display is updated Using the or push button select 0 5 0 5 seco...

Page 26: ...h buttons The button will transfer control to the second digit which controls the step response and should be set to 0 in the same way Finally enter the selection and return to the FiLtEr prompt in th...

Page 27: ...s unauthorised access to the configuration menu Using the or button select CodE from the configuration menu and press which will reveal 0000 with the first digit flashing This example requires the sec...

Page 28: ...ion change Re enable both alarms See 9 3 4 8 2 Fault finding after commissioning EN 60079 17 Electrical installations inspection and maintenance permits live maintenance in Zone 2 if a risk analysis d...

Page 29: ...your local BEKA agent It is helpful if a brief description of the fault symptoms is provided 8 6 Customer comments BEKA associates is always pleased to receive comments from customers about our produ...

Page 30: ...with factory fitted dual isolated solid state single pole alarm outputs that may be independently configured Each may be configured as a speed or run time alarm with a high or low function having a no...

Page 31: ...the means of isolation is essential for their effective use It is also necessary to ensure that the maintenance procedure makes sure that unauthorised re closure of the switches does not occur It is n...

Page 32: ...t and the alarm output being activated See section 9 3 10 5IL Alarm silence time Defines the time that the alarm output remains in the non alarm condition following acceptance of an alarm See section...

Page 33: ...to return to the alarm sub menu 9 3 5 Type of alarm tYPE Alarm 1 and Alarm 2 are totally independent both may be speed or run time alarms or one may be conditioned for speed and the other for run time...

Page 34: ...ggle the contact status between no and nC when set as required press the button to return to the no nC prompt in the alarm sub menu 9 3 9 Hysteresis H5tr Hysteresis is only available on speed alarms s...

Page 35: ...l adjustment is provided by a separate security access code Using the or push button select AC5P from the configuration menu and press to reach the enable function EnbL Pressing will reveal the existi...

Page 36: ...the or push button and the button to transfer control to the next digit When the required setpoint has been entered pressing will return the display to the 5P1x or 5P2x prompt from which the other se...

Page 37: ...ed and contain a means of isolation These two requirements may be satisfied by installing DIN rail mounted terminals incorporating a fuse which can easily be opened as shown in Fig 2 to provide isolat...

Page 38: ...g the or push button select 4 000 in the 4 20mA output sub menu and press to reveal the existing speed display with one digit flashing The required display can be entered using the or push button to a...

Page 39: ...cate indicated by the certificate number s X suffix These state that the BA314NG Tachometer should be a Supplied from a limited energy circuit with output parameters in normal operation equal to or le...

Page 40: ...ical to the ATEX safety parameters described in the main section and in Appendix 1 of this manual The IECEx certificate may be downloaded from the BEKA associates or the IECEx website or may be reques...

Page 41: ...T5 Gc Zone 22 AEx ic tc IIIC T80 C Dc Ta 40 C to 60 C cETL codes for Canada Ex nA ic IIC T5 Gc Ex n IIC T5 Gc Ex ic tc IIIC T80 C Dc Class III Div 2 Class II Div 2 Gp F G Ta 40 C to 60 C The ETL and...

Reviews: