background image

9

6.1

Summary of configuration functions

This  section  summarises  each  of  the  main
configuration  functions  and  includes  a  cross
reference   to   a   more   detailed   description.     Fig   7
illustrates the location of each function within the
configuration  menu.    The  lineariser  and  the
optional  factory  fitted  alarms  are  described
separately in sections 7 and 9.3 of this manual.

   Display

      Summary of function              

FunC

Indicator function

Defines  the  relationship  between  the
4/20mA input  current  and  the  indicator
display.  May be set to:

5td

Standard linear relationship

root

Square root extraction

Lin

16 segment adjustable
lineariser – see section 7.

See section 6.2

rESn

Display resolution

Defines  the  resolution  of  the  least
significant  display  digit.    May  be  set  to

1

2

5

 or 

10

 digits.

See section 6.3

dP

Decimal point

Positions  a  dummy  decimal  point
between any of the digits or turns it off.
See section 6.4

CAL

Calibration of the digital display using
an external current source.

Enables  the  zero  and  span  of  the
indicator to be adjusted using an external
current  source  such  as  a  calibrator.
When  used  with  an  accurate  traceable
current  source  this  is  the  preferred
method of calibration.
See section 6.5

5Et

Calibration  of  display  using  internal
references.

Enables  the  zero  and  span  of  the
indicator to be adjusted without the need
for  an  accurate  input  current  or
disconnection from the 4/20mA loop. See
section 6.6

bAr

Bargraph format and calibration

Only the BA324NG has a bargraph

The bargraph may be conditioned to start
from left, right or centre of the display, or
it  may  be  disabled.    When  optional
alarms are fitted it can also display both
alarm setpoints and the measured value. 
The bargraph may be calibrated to start
and  finish  at  any  value  within  the
indicator’s calibrated digital display.
See section 6.7

   Display

      Summary of function              

C--P

Function of 

(

 push button

The  indicator  may  be  configured  to
display the input current in milliamps, or
the input current as a percentage of the
4/20mA input when the 

(

 push button is

operated in the display mode.
See section 6.8

tArE

Tare function

 When enabled the tare function sets the
indicator  display  to  zero  when  the  

)

push  button  is  operated  in  the  display
mode.
See section 6.9

CodE

Security code

Defines  a  four  digit  numeric  code  that
must  be  entered  to  gain  access  to  the
configuration menu.   Default code  

0000

disables this security function and allows
unrestricted  access  to  all  conditioning
functions.
See section 6.10

r5Et

Reset

Contains  two sub-functions,  

ConF

 which

returns  the  indicator  to  the  default
conditions shown in section 6.0 and 

LtAb

which returns the lineariser to the default
conditions  shown  in  section  7.3.      To
prevent accidental use both resets must
be  confirmed  by  entering  

5urE

 before

they will be executed.
See section 6.11

Summary of Contents for BA304NG

Page 1: ...Issue 4 1st November 2017 BA304NG BA324NG Ex nA Ex tc loop powered field mounting indicators Issue 4...

Page 2: ...tor 7 3 Under and over range 7 4 Lineariser default configuration 8 Maintenance 8 1 Fault finding during commissioning 8 2 Fault finding after commissioning 8 3 Servicing 8 4 Routine maintenance 8 5 G...

Page 3: ...Notified Body Intertek Testing and Certification Ltd which has been used to confirm compliance with the European ATEX Directive 2014 34 EU For international applications both models have IECEx certifi...

Page 4: ...and dust certification This section of the instruction manual describes ATEX gas certification ATEX dust certification is described in Appendix 1 and IECEx approvals are described in Appendix 2 3 1 A...

Page 5: ...to display the measured variable or control signal in engineering units The indicators are transparent to HART signals There are four design requirements 1 The current flowing in the 4 20mA loop must...

Page 6: ...on the rear of the indicator assembly as shown in Fig 5 If the field wiring is to be terminated prior to the installation of the indicator assembly a back box terminal assembly which includes a 4 20m...

Page 7: ...ors are supplied with a printed scale card showing the requested units of measurement and tag information If this information is not supplied when the indicator is ordered a blank scale card will be f...

Page 8: ...r this prompt allowing each digit of the code to be entered using the and push buttons and the button to move control to the next digit When the correct four digit code has been entered pressing will...

Page 9: ...djusted without the need for an accurate input current or disconnection from the 4 20mA loop See section 6 6 bAr Bargraph format and calibration Only the BA324NG has a bargraph The bargraph may be con...

Page 10: ...10...

Page 11: ...not required it should be positioned beyond the most or least significant digit When positioned as required press the button to enter the selection and return to the configuration menu 6 5 Calibration...

Page 12: ...cator s digital display at which the bargraph starts is defined by the bArLo sub function which is selected by pressing the button followed by the button which will reveal the current indicator displa...

Page 13: ...ions shown in sections 6 0 and 7 3 To reset the indicator or lineariser select r5Et from the configuration menu and press the indicator will display one of the reset options ConF or LtAb ConF Resets t...

Page 14: ...ows direct calibration of the lineariser with an external current source and is the preferred method when traceability is required If the exact system non linearity is unknown this method also allows...

Page 15: ...15...

Page 16: ...ng break point 7 1 1 Example Adding break points to a new indicator When adding break points to a new indicator or to a lineariser following resetting to the factory defaults using the LtAb function d...

Page 17: ...n by pressing the button The corresponding indicator display at each of the break points can now be defined using the di5P sub function Using the and buttons select the di5P sub function and press whi...

Page 18: ...o risk of a flammable atmospjher being present If an indicator fails after it has been functioning correctly follow the procedure shown in section 8 1 If this does not reveal the cause of the fault it...

Page 19: ...haracters 5mm high 9 2 Display backlight The BA304NG and BA324NG loop powered indicators can be supplied with a factory fitted backlight that may be loop or separately powered When loop powered the ba...

Page 20: ...s an emergency shut down system Both models can be supplied with factory fitted dual solid state single pole alarm outputs Each alarm output may be independently conditioned as a high or low alarm wit...

Page 21: ...lication in which a BA304NG or BA324NG indicator is displaying the output from a 2 wire transmitter in Zone 2 Alarm 1 is switching a solenoid valve in Zone 2 and alarm 2 is switching a sounder located...

Page 22: ...tpoint The alarm is activated when the indicator display equals the setpoint See section 9 3 5 Hi Lo Alarm function Defines the alarm function as High or Low See section 9 3 6 no nC Normally open or n...

Page 23: ...23...

Page 24: ...e for fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and pr...

Page 25: ...t having access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This dir...

Page 26: ...plus the displayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 18 Fig 18 Displayed value and setpoints on bargraph T...

Page 27: ...supplied with each panel mounting kit 9 6 Back box terminal assembly If it is necessary to terminate the field wiring prior to the installation of the indicator display a back box terminal assembly is...

Page 28: ...o be used for Zone 22 combustible dust applications The indicator should be located where the minimum amount of dust will accumulate on the enclosure and the installation should be as specified in thi...

Page 29: ...USA and Canada A3 0 cETL Mark For installations in the USA and Canada the BA304NG and BA324NG indicators have ETL and cETL Ex nA and Ex tc approval Control Number 4008610 Copies of the Authorisation...

Reviews: