background image

BEAR | 2017 | 

page 20

13. Maintenance of the water distiller without 

built in tank

 

13.1.  The routine maintenance during the operation period and periodical maintenance 

should be provided for the water distiller:

13.2.  Daily check visually the self-cleaning filter for clogging. In the case of contamination, 

wash it with water. 

13.3.  In case of a decrease in water still performance, significant steam emission from the 

water still, you should remove salt-contained deposits from the heat exchanger and 

the orifice of the nozzle 9. To do this, unscrew the ball valve and the water intake pipe 

7 (Figure 5), unscrew the nozzle 9, tighten the fitting without nozzle. Start the water 

distiller for 30 seconds in the working mode, then turn off the water distiller and insert 

the nozzle in the reverse order having cleaned the nozzle orifice with a bodkin.

13.4.  Once a month (or more often if necessary), the evaporation chamber and electric 

heaters should be cleaned of scale using citric acid solution (50 g per liter). To do this, it 

is necessary to fill solution into the evaporator.  Fill evaporator with water, add cleaning 

solution for descaling. 

Close the raw water supply valve.

 Bring to a boil and boil for 

10-20 minutes. Then, after 60 minutes, drain the solution 

13.5.  Rinse with water and pour the solution of baking soda for 5 minutes (10 g per liter), drain 

and finally rinse with plenty of water 3-4 times. If necessary, treatment can be repeated. 

13.6.  In case of poor quality of raw and, consequently, produced water, must be completely 

drained from the water still and the inner surfaces of condensation and evaporation 

chambers should be thoroughly cleaned by flushing with solution of baking soda (10 g 

per liter) followed by wiping (especially along the contour of bottoms of the evaporator 

and collector) with a lint-free cloth.

14. Transportation

14.1.   The water distillers should be transported in covered vehicles (railway cars, containers, 

covered motor vehicles, holds, etc.), and the containers with the water distillers should 

be attached securely in order to prevent any possible displacement.

15. Storage

15.1.  The packaged water distiller should be stored indoors at an ambient temperature of   

minus 50˚С up to plus 40˚С and a relative humidity of up to 98 % at +25˚С.

15.2.  The air in the premises should not contain such impurities as corrosive vapors and 

gases.

Summary of Contents for A 1104

Page 1: ...Water Distillers OPERTION MANUAL A 1104 A 1110 A 1125 A 1204 A 1210...

Page 2: ...ke modifications to the design of the water distiller Be careful water distiller surfaces and distillate can be extremely hot while operating Use only original accessories and parts heaters valves con...

Page 3: ...controlled by personnel all the time while operating Make sure that distillate receptacle is big enough and not full before starting operation 2 General information 2 1 Water distillers BEAR B 1104 BE...

Page 4: ...Due to heaviness of BEAR B 1110 and BEAR B 1125 those models should be unpacked by 2 persons For BEAR B 1110 and BEAR B 1125 the easiest way is to disassemble the wooden crate and then remove the pal...

Page 5: ...n hose inlet hose length 2 meters 1 Water connection hose drain hose length 1 5 meters 16x25 mm 2 Fitting with male thread 12 mm 2 Silicone tube for constant distillate withdrawal 15x3 mm length 1 met...

Page 6: ...1 Fitting with male thread 12 mm 1 Metal water connection hose inlet hose length 1 5 meters 1 Water connection hose drain hose length 1 5 meters 2 Silicone tube for constant distillate withdrawal 16x2...

Page 7: ...ler design the structure component view is schematic not taking into considera tion features of the water stills Figure 2 shows the fitting grounding terminal connection diagram 4 Water distiller with...

Page 8: ...changer inlet 16 Evaporator inlet 17 Heater 18 Equalizer drain pipe 19 Heat exchanger 20 Condenser drain pipe 21 Raw water supply valve 22 Collector distillate level pipe 23 Evaporator top level senso...

Page 9: ...vaporator 2 steam forms then separated and enters in the condenser 6 via the channel 5 The separator consists of the steam floe splitter 3 and the nozzle 4 located inside the channel 5 4 4 Steam conde...

Page 10: ...utomatic device switches the heater OFF at the same time the switch led HEATER turns off 5 5 The Low water thermostat sensor 10 Figure 1 ensures an additional emergency shut down of the water still in...

Page 11: ...sideration features of the water stills 6 Water distiller without built in tank design and principle of operation 1 Housing 2 POWER switch 3 Electric heater s 4 Low water thermostatic sensor 5 Distill...

Page 12: ...ld be used to collect distillate 7 Before installation Installation of the equipment must be done only by qualified personnel and only after reading this operating manual 7 1 The water distillers are...

Page 13: ...IGH WATER LEVEL red LED will light up 8 6 Install gaskets in the filling hose nuts if they ar not pre installed at the factory 8 7 Connect the filling hose to the fitting 25 Figure 2 8 8 Connect the f...

Page 14: ...14 Connect the water still grounding terminal to the PG bus with flexible copper wire at least 6 mm2 in cross section Connect the external circuits to the power supply network in accordance with Figur...

Page 15: ...re Equalizer is about 70 80 degrees Celsius 9 5 Install water distiller on the table using the level gauge and avoiding slants 9 6 Attach the silicone tube L 0 5 m to the distillate drain pipe 5 9 7 C...

Page 16: ...s an overflow of water and turn off the water still and the HIGH WATER LEVEL LED will be lit 10 7 To start the water distiller push the POWER button located on the front panel and this button should b...

Page 17: ...d down and then push the lock button located on the thermal switch housing Pos 13 Figure 5 11 6 To start the water still push the POWER button located on the front panel and this but ton should be lit...

Page 18: ...splitter 3 Figure 1 12 6 6 Remove the steam flow splitter 3 12 6 7 Flush with water the walls and bottom of the evaporator to remove salt con taining sediments You can use the filling hose for supplyi...

Page 19: ...12 6 28 Make sure that the sensors 9 23 Figure 1 is rigidly fixed and do not contact the walls 12 6 29 Assembly the water still performing steps 4 3 in reverse order 12 6 30 The water still can be use...

Page 20: ...evaporator with water add cleaning solution for descaling Close the raw water supply valve Bring to a boil and boil for 10 20 minutes Then after 60 minutes drain the solution 13 5 Rinse with water an...

Page 21: ...ARS Guarantee mentioned above is limited for the heaters for 6 months 16 2 Service Maintenance of the distillers should be performed by qualified personnel Distiller serial number_____________________...

Page 22: ...Contact Info Location 5115 Douglas Fir Road Suite G Calabasas California 91302 USA Phone 424 259 0816 E mail info bearinnovation com...

Reviews: