the mains using the plug for 1.5kW versions of BEAR B 1104 and BEAR B 1204 and for
the rest of the models using mains input circuit breaker (Figure 4) to be installed by the
customer and intended for protection against overloads and short-circuits of electrical
circuits.
8.13. The circuit breaker is not supplied, should be purchased by the customer and selected
based on the calculation of the In+10% overcurrent tripping setting (where In is a rated
current).
8.14. Connect the water still grounding terminal to the PG bus with flexible copper wire at
least 6 mm2 in cross section. Connect the external circuits to the power supply network
in accordance with Figure 4 and 5
Input circuit breaker
Five core cable
Grounding
Terminal
yellow-green
HOUSING
BA
blue
L1
L2
L3
N
PG (Ground)
Input circuit breaker
Socket and plug
with a grounding
contact
Grounding
Terminal
HOUSING
BA
L1
N
PG (Ground)
Figure 4. Wiring diagram for the connection of
external circuits of water distillers to the single
phased 120 VAC electric network
Figure 5. Wiring diagram for the connection of
external circuits of water distillers to the 3-Y 220
VAC electric network