BDDF00171 Issue 4
35/39
BD Alaris™ MRI Capsule
Electromagnetic Emissions
Electromagnetic Emissions
Guidance and Manufacturer’s Declaration
– Electromagnetic Emissions
The MRI Capsule is intended for use in the electromagnetic environment specified below .
The customer or the user of the MRI Capsule should assure that it is used in such an environment .
Emissions Test
Compliance
Electromagnetic Environment – Guidance
CISPR 11
RF Emissions
Group 1
The MRI Capsule uses RF energy only for its internal
function in the normal product offering . Therefore, its RF
emissions are very low and are not likely to cause any
interface in nearby electronic equipment .
CISPR 11
RF Emissions
Class A system when used in
conjunction with pumps
Class B in stand alone
operation
The MRI Capsule is suitable for use in all establishments,
other than domestic, and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes .
EN 61000-3-2
Harmonic Emissions
Class A
EN 61000-3-3
Voltage Fluctuations,
Flicker Emissions
Complies