BDDF00171 Issue 4
4/39
BD Alaris™ MRI Capsule
Introduction
Intended Positioning of the MRI Capsule
The MRI Capsule incorporates a Magnetic Indicator that is used to safely position the MRI Capsule in relation to the MRI
Scanner . The MRI Capsule should be positioned as near to the patient as possible, ensuring the Magnetic Indicator is
illuminated Green which indicates a magnetic field density equal to or less than 20mT (millitesla) .
Intended user population
The MRI Capsule and its compatible Pumps can only be used by trained and qualified medical staff within the MRI
environment .
Intended patient population
The MRI Capsule can only be used on one patient at a time, and it serves the same population of patients as the
compatible Alaris™ Infusion Pumps .
Contraindications
The MRI Capsule is not designed to be used in Home Care environments . Please refer to the 'Product Specifications'
section for more information .
Incompatible Alaris™ Infusion Pumps and Accessories for MRI Capsule
Pump Model
Model Code
Alaris™ GW Volumetric Pump
All variants
Alaris™ GW 800 Volumetric Pump
All variants
Alaris™ GP Volumetric Pump (all variants)
All variants
Flow Sensor
180
180A