BDDF00171 Issue 4
27/39
BD Alaris™ MRI Capsule
Product Specifications
Product Specifications
Electrical
Protection Against Electrical Shock:
Class I
Supply Voltage:
115–230V, ~50–60Hz
Rating:
460VA (Maximum)
Fuses:
2 × 4A/H Time-Lag T
Consult Qualified Service Personnel for replacement of fuses .
AC Power Outlets:
MRI Capsule Tile: 115–230V, ~50–60Hz, 60VA
Protection against fluid ingress:
IPX2 – Protection against water droplets deflected up to 15° from vertical .
Battery
MRI Capsule Battery
Type:
Nickel Metal Hydride
Charge time:
With no Pumps attached: Up to 8 hours to 95% charge time based on battery life cycle .
With Pumps attached: Up to 16 hours to 95% charge time based on battery life cycle and a 6 Pump load .
Operating time:
60 minutes
Magnetic Indicator Battery
Type:
Primary Alkaline Battery (Zn/MnO
2
)
Operating time:
Approximately 2 years, under normal conditions
Environmental
Operating
Transport and Storage
Temperature:
+5°C to +40°C
−20°C to +50°C
Humidity:
20% to 90%
20% to 90%
Atmospheric Pressure:
70 to 106kPa
50 to 106kPa
Mode of Operation
Continuous Operation