BD Alaris™ MRI Capsule
The MRI Capsule in the MRI Environment
Removing a Pump
1 . Unlock and open the MRI Capsule door .
2 . Holding the Pump with both hands, push the release lever on the right hand side of the Pump backwards .
3 . Keeping the lever pushed back, pull the Pump horizontally towards you .
4 . Slide the Infusion Sets away from the grooves
5 . Check that the green LED indicator on the MRI Capsule tile is not extinguished after removal of the Pump with AC
Power present on the MRI Capsule .
Note: When removing an Alaris™ VP Plus Guardrails™ Volumetric Pump from the top left MRI Tile follow the best
practice specified below .
w
If the indicator in the MRI Capsule tile is
not
illuminated when no infusion Pump is attached to the MRI
Capsule tile, (with AC Power present) suspect a fault with the MRI Capsule tile. Remove the MRI Capsule
from service and contact Qualified Service Personnel.
Best practice for docking and undocking Alaris™ VP Plus Guardrails™ Volumetric Pumps
Hand positioning for docking Pump on top left MRI Tile
Hand positioning for undocking Pump on top left MRI Tile
Positioning of hand, as shown opposite, when
handling the Pump in the MRI Capsule could
result in user harm.