BDDF00171 Issue 4
21/39
BD Alaris™ MRI Capsule
Operating Precautions
• If the MRI Capsule is moved to an area where the Magnetic Indicator is flashing red, and
the Pumps do not alarm, the user must move the MRI Capsule back to a position where
the Magnetic Indicator flashes green . The infusions can continue without interruption . The
Pumps must be checked by qualified service personnel when the infusions are complete .
• If the MRI Capsule is moved in to an area where the Magnetic Indicator is flashing red
momentarily, and Pumps do not alarm, user must move the MRI Capsule back into an area
where the Magnetic Indicator is flashing green and continue infusions without interruption .