BD Alaris™ MRI Capsule
Electromagnetic Emissions
Recommended Separation Distances for between portable and mobile RF communications
equipment and the MRI Capsule
The MRI Capsule is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled .
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any part of the MRI Capsule, including cables specified by the manufacturer .
Otherwise, degradation of the performance of the MRI Capsule could result .
The MRI Capsule should not be used next to other equipment . If adjacent use is necessary, the MRI Capsule should be
observed to verify normal operation in the configuration in which it will be used .
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment in accordance to IEC
60601-1-2:2014:
Test frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380 - 390
TETRA 400
Pulse
Modulation
18 Hz
1 .8
0 .3
27
450
430 - 470
GMRS 460,
FRS 460
Pulse
Modulation
18 Hz
2
0 .3
28
710
745
780
704 - 787
LTE Band 13,
17
Pulse
Modulation
217 Hz
0 .2
0 .3
9
810
870
930
800 - 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
Modulation
18 Hz
2
0 .3
28
1720
1845
1970
1700 - 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
Modulation
217 Hz
2
0 .3
28
2450
2400 - 2570
Bluetooth,
WLAN,
802 .11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
Modulation
217 Hz
2
0 .3
28
5240
5500
5785
5100 - 5800
WLAN 802 .11
a/n
Pulse
Modulation
217 Hz
0 .2
0 .3
9