If the label stock contains preprinted information, position the sensor
over an unmarked label area.
Three Media Sensor readings are required:
Empty:
Nothing in the sensor.
Mark or Gap: Only the backing material, notch, or reflective mark in the sensor.
Paper:
The label (with the backing material attached) in the sensor.
To perform a Standard Calibration:
Step
Operator Action
Displayed Message
Comment
1
Turn ‘On’ the printer.
"
Wait until
‘Ready’ is
displayed.
2
Press the
key to enter
Menu Mode.
Raise the printhead for
visual access to the Media
Sensor and load label
media.
"
See Section
4.0.2 for
details.
3
Press the
key to
enter the Media Settings
menu.
"
"
See Section
4.1.1 for
details.
4
Press the key.
"
See Section
4.1.1 for
details.
5
Press the key. Use
the key to scroll
to the desired sensor type
and press the
key
to enable the selection.
J"
See Section
4.1.1 for
details.
The enabled
selection will
be indicated
with an *
.
(Continued next page)
Summary of Contents for 4206
Page 1: ...Meto mi 4 Label and Tag Printer Extended Operating Manual B a r c o d e S y s t e m s ...
Page 5: ......
Page 7: ...ii ...
Page 9: ...iv ...
Page 11: ...vi ...
Page 12: ...vii ...
Page 13: ......
Page 17: ...4 mi 4 Series ...
Page 20: ...mi 4 Series 7 ...
Page 32: ...mi 4 Series 19 ...
Page 44: ...mi 4 Series 31 ...
Page 50: ...mi 4 Series 37 Continued next page ...
Page 56: ...mi 4 Series 43 ...
Page 58: ...mi 4 Series 45 ...
Page 62: ...mi 4 Series 49 ...
Page 74: ...mi 4 Series 61 ...
Page 78: ...mi 4 Series 65 Continued next page ...
Page 80: ...mi 4 Series 67 ...
Page 82: ...mi 4 Series 69 ...
Page 84: ...mi 4 Series 71 ...
Page 86: ...mi 4 Series 73 ...
Page 111: ...98 mi 4 Series ...
Page 114: ...mi 4 Series 101 ...
Page 139: ...126 mi 4 Series Continued next page ...
Page 142: ...mi 4 Series 129 ...
Page 144: ...mi 4 Series 131 ...