3 Disassembly / Assembly
3 - 32
Infusomat® compact
plus
P 1.0
EN
For internal use, only
Assembly
1. If the housing, bottom part has been replaced, mount the lock
screw (
) of the old housing to the new
housing from below, or use a new lock screw from the Small
parts ICP.
2. Tighten the screws of the power supply unit with a torque of
0.85 Nm ± 0.1 Nm.
CHECKS AFTER REPAIR
Procedure
Always perform the check after repair whenever any repair work
has been done on the unit (
).