4 Check after Repair
Infusomat® compact
plus
P 1.0
4 - 11
EN
For internal use, only
4. After the 3 minutes have elapsed, read off the corresponding
value for P
max
(observe for 30 seconds) on the pressure gauge
and compare with the specifications for P
max
in the checklist.
5. Document value (P
max
). Do not reduce pressure.
T7 SHK
1. Open the pump door.
2. Check if opening the door has caused closing of the SHK
, left).
3. Move the pump door towards the closed position as far as
possible keeping the door opener in closed position.
The second clamp opens (see
4. Read off the P
SHK
value on the pressure gauge after min.
10 seconds and compare with the specifications for P
SHK
in the
checklist.
5. Document value (P
SHK
). Release the pump door.
NOTICE
If the pressure values are not reached, the mechanical pressure
can be adjusted without opening the unit
.
Fig. 4 - 5
Fig. 4 - 6
SHK closed (left) and opened (right)
Fig. 4 - 7
Second clamp opened (right)