3 Disassembly / Assembly
Infusomat® compact
plus
P 1.0
3 - 5
EN
For internal use, only
3.2
GASKET FOR RUBBER CONNECTOR
Designation Ord.
No.
Gasket for rubber connector CP. . . . . . . . . . . . . . . . . . . . .34522010
Screws
Disassembly
1. Unscrew the four screws (
) using a 10IP TORX
plus screwdriver.
2. Remove the gasket (
).
Assembly
✓
Power supply is installed in the housing, bottom part.
1. Attach the gasket with the correct orientation.
2. Tighten the screws with a torque of 0.75 Nm ± 0.1 Nm.
Fig. 3 - 3
1 Screw DELTA PT 30x9 WN 5454
2 Gasket
NOTICE
Use a new gasket whenever the gasket has been removed.