Infusomat® compact
plus
P 1.0
2 - 1
EN
For internal use, only
2 Unit Diagnosis / Adjustments
ALARMS AND ERROR MESSAGES
The alarms of the Infusomat® compact
plus
P are grouped into
different categories. The categories are listed below in order of
increasing importance.
■
Advisory alarm
A message is output if incorrect entries are made (e.g.
“Maximum reached“, “The parameter cannot be changed“) and
an audible signal is output.
■
Reminder alarm
A reminder alarm is triggered if the device is not operated for
two minutes after an incomplete entry or operator action on
the device.
■
Pre-alarm
Pre-alarms are triggered several minutes (depending on the
service settings) before the operating alarms.
■
Operating alarm
If an operating alarm is output, infusion is stopped. An audible
signal is output, the red LED flashes and a staff call is triggered.
The message “Alarm” and the cause of the alarm appear on the
display.
■
Device alarm
If a device alarm is output, infusion is stopped immediately. If
the device is restarted and the alarm is displayed again,
servicing of the unit is required.
The most important alarms and error codes as well as their
meaning and possible fault clearance are specified in the following
sections.
LOG FILES
The device logs activity including security relevant events in a ring-
buffer.
This buffer is protected against manipulation, i.e. it is not possible
to add, change, or delete single entries. Each entry includes date,
time and event description. The files are stored in the device
memory and can be retrieved using the Service Tool (
see
"compact
plus
Service Tool" Instructions for Use
).
The log files are not automatically analyzed.
PRE-ALARMS AND OPERATING ALARMS
See Instructions for Use.