3 Disassembly / Assembly
3 - 10
Infusomat® compact
plus
P 1.0
EN
For internal use, only
3.6
SECOND CLAMP
Designation Ord.
No.
Second clamp ICPP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34522286
Screw and cover cap
Disassembly
✓
Pump door is open.
1. Pierce the cover cap (
) with a small screw-
driver or twist gimlet and remove it.
) using a 10IP TORX plus
screwdriver.
3. Slightly lift the second clamp (
) toward the
top and remove it to the side.
Assembly
✓
Unit is closed.
1. Tighten the screw with a torque of 0.65 Nm ± 0.1 Nm (first and
second screwing).
Fig. 3 - 10
1 Cover cap
2 Screw DELTA PT 30x10 WN 5452
3 Second clamp
NOTICE
If the second clamp ICPP has to be replaced, the transport
bracket must be removed from the new spare part before it is
mounted to the pump.