![bbraun 8717070 Service Manual Download Page 86](http://html1.mh-extra.com/html/bbraun/8717070/8717070_service-manual_2714001086.webp)
7 Procedural Instructions on the TSC
7 - 2
Infusomat® compact
plus
P 1.0
EN
For internal use, only
ELECTRICAL SAFETY
ACCORDING TO IEC 62353
The values to be measured for electrical safety are listed in the TSC
(
see "Technical Safety Check (TSC)"
Equipment leakage current (alternative method)
1. Measure between mains input and pin 3 (GND) of accessory
connector (
).
For safer measurement, use the Measuring Adapter Accessory
Connector CP.
FUNCTIONAL INSPECTION
Mechanical inspection
1. Fit the unit to be tested on top of another compact
plus
device
and check that the lock functions correctly.
2. Fit the unit to be tested underneath another compact
plus
device
and check that the lock functions correctly.
3. Rotate the pole clamp in all directions (4x90°) and check for
correct latching.
4. Check that the rotary knob of the pole clamp can be turned
clockwise and counterclockwise.
Operational test
1. Connect the unit to the power supply without switching it on
and check the status display:
a) Battery capacity
b) Mains operation symbol
2. Switch the unit on and check the following details:
a) Self-test
b) Display of software version
c) Audible alarm
one high sound from piezo buzzer
one deep sound from loudspeaker
d) Visual alarm
The green LED flashes once. The red LED flashes twice.
Fig. 7 - 1
1 Pin 3 of accessory connector
▲
Top of unit
▼
Bottom of unit