PD-TaD 62
Product information
822-187-3
19 / 114
No.
Element
Function
1
Mains voltage
connection
Is used to connect the Power Box to the mains voltage
When the mains voltage is connected, the LED next to the mains voltage
connection lights up.
2
Device protection fuse
Used to protect the Power Box and connected system components
Type: T 16 AH
3
Laptop
port
Is used to connect the laptop
The laptop is used for data transmission and is connected to the Power Box
via the Ethernet cable (included in the standard delivery).
4
PD Detector
port
Is used to connect the PD measuring unit of the PD-TaD 62
The PD measurement unit is connected to the Power Box via a PoE cable.
The PoE cable is used for the power supply of the PD-TaD 62 and for data
transmission.
5
Ports for system power
supply
Are used to connect the VLF HV generator and the laptop to the power
supply on the Power Box
Max. current 16 A
3.5
HF filter
The HF filter is used to minimise external influences on the measurement results. The HF
filter is supplied in the PD-TaD 62 transport case.