SEC2000 Operating Manual
SEC2000 Spectra System
www.als-japan.com
65
⑥
When the external TTL changes from 5 V to 0 V, the spectrum measurement is
begun, and the spectrum and the monitor screens start to move (Fig. 8-3-6).
<Fig. 8-3-6 >
⑦
To stop the monitor, you can configure the “
Duration
” or click the “
STOP
” icon in
the Mornotor1 screen.