SEC2000 Operating Manual
SEC2000 Spectra System
www.als-japan.com
33
<Fig 5-4-1>
①
First, the light source and detector of SEC2000 are set as shown in Fig. 5-4-1.
②
The power supply is switched on and turned ON the UV lamp and halogen lamp.
Wait for 20 to 30 minutes until the lamps are stabilized.
③
Open the software and click “
Go
” icon, the spectrum signal will appear. On Scope
Mode, put the reference sample into the cuvette holder, and open the light source
shutter fully.
④
Adjust the Integration Time to control the maximum value of the Y axis of the graph
(less than 16,000).
⑤
By clicking the “
Reference
” icon in the program, the spectrum is obtained at the
maximum light intensity. Save the reference spectrum by “
File
”/”
Save
”/
“
Reference
”.
⑥
Close the light source shutter (Do not close the power switch), and click the “
Dark
”
icon in the program to obtain a spectrum under the minimum conditions without
light. Save the dark spectrum by “
File
”/”
Save
”/”
Dark
”.
⑦
Put the sample into the cuvette holder, and open the light source shutter fully. Click
the “
Absorbance Mode
”/”
Transmittance Mode
” icon to change the Mode. Store
the spectrum by “
File
”/”
Save
”/”
Processed
” or “
Sample
” to save an absorbance or
transmittance spectrum.
Notes
When absorbance/transmittance is saved by “
File
”/”
Save
”/”
Processed
”, the data of
absorbance/transmittance is displayed only absorbance/transmittance mode. However,
if absorbance/transmittance is saved “
File
”/”
Save
”/“
Sample
”, the saved data is
convertible for other mode.