SEC2000 Operating Manual
SEC2000 Spectra System
www.als-japan.com
25
④
Overlay
-
Spectrum
-
Time Series
-
Add Overlay 1 ~ 10
Display the data for each spectral window (1 ~ 10).
-
Clear
Clear selected overlay data from the spectral window.
-
Add Overlays (Multiple Selection)
Display several data at the same time.
-
Clear All
Clear all overlay data from the spectral window.
⑤
Spectrum
-
Config. Spectrometer
Confirm the parameters of spectrometers for
wavelength
calibration.
<Fig. 4-3-8>
-
Reset Spectrometer
This function cannot be used for user.
-
Calib. Spectrometer
This function cannot be used for user.
-
Store Dark
Save the spectrum measurement result as dark spectrum to
memory and display on overlay 1.
-
Store Reference
Save the spectrum measurement result as reference
spectrum to memory and display on overlay 2.
-
Snapshot
Halt data acquisition and take a snapshot of activity in the
spectral window.
-
Scope Mode
Present the data in scope mode.
-
Scope Mode Minus Dark
Switch the current spectral window into scope mode
and subtract the stored dark spectrum.
-
Absorbance Mode
Display data in absorbance mode. Please save the Dark
and Reference spectra before using this function.
-
Transmittance Mode
Display data in transmittance mode. Please save the