Display information
40
ambient light value. Use Enter
or click the control
wheel again to store the value.
• However, if you wish to adapt the ambient light
conditions to the definition of the selected reading room,
dim the light in the room so that the measured ambient
light becomes lower that the maximum ambient light.
• Then use Enter
or click the control wheel again to
execute the Measure ALC Value function once more. As a
result the actually measured ambient light value will be
copied as Measured ALC Value. Finally use Enter
or
click the control wheel again to store the value.