Display information
27
Reading room definition submenu
Name
Description
Measured ambient light
Indicates the ambient light actually
measured by the ambient light sensor
on the front of the display. This is a
read-only value.
Averaged ambient light
Shows the average of the measured
ambient light since the display was
switched on. This is a read-only value.
Reading room
Indicates the reading room type you are
editing. You can select another room to
edit by using the Up
or Down
touch keys or rotating the control
wheel.
Max. ambient light
This indicates the maximum ambient
light that corresponds to the selected
reading room.
If the measured ambient light is higher
than the value entered here, you should
take measures to darken the room or
select another reading room type.
Preset corr. value
This value is taken into account in the
calculation of the DICOM DF if Continu-
ous DICOM ALC is switched off and
DICOM Offset is set to “Preset”.
To each reading room type corresponds
another preset value.