Display information
25
Averaged ambient light
Shows the average of the measured
ambient light since the display was
switched on.
When you execute the function “Mea-
sure ALC Value”, this value will be
stored as the Measured ALC correction
value for the selected reading room,
unless this value is higher than the
maximum ambient light determined for
the selected reading room.
Correction value
Shows the ambient light correction
value that is taken into account in the
calculation of the display function.
Cont. DICOM ALC
Allows to switch continuous DICOM ALC
on/off.
When switched on, the DICOM DF is
recalculated
continuously
taking the
averaged ambient light into account.
When switched off, the DICOM DF is
recalculated
only
at the moment the
DICOM Offset is changed (see below).
This function is available only when the
DICOM DF is selected.
Name
Description