Display information
21
backlight value). The message will not disappear until the display is
powered off.
LCM Temperature Sensor Off Line
This is a serious error message. You should power off the display
and call for service.
The warning means the LCD panel temperature sensor has lost
communication. As a result the display switches to a safe operating
mode (with decreased backlight value). The message will not
disappear until the display is powered off.
On-screen display (OSD)
Luminance and color menu
Luminance and color menu
Name
Description
Measured luminance
Indicates the actual luminance mea-
sured by the internal sensor. This is a
read-only value. It is expressed in
cd/m2.
Luminance target
Allows to manually adjust the lumi-
nance target. See note below.
Color target
Allows to select from a list of factory-
defined and user-defined color targets.