BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER
24
2.4 Dip the electrode into the next calibration solution, the meter automatically shows current calibration standard (e.g. 12.88mS/cm).
2.5
Press ENTER key to confirm, “Calibration” indicator
begins flashing again.
2.6 Wait for the measured value to stabilize, the display shows
“CAL3”
, the meter prompts you to continue with third point calibration.
2.7 Repeat steps 2.3 to 2.5
above until the display shows “END”
, the meter returns to measurement mode automatically. Calibration is
completed.
MANUAL CALIBRATION:
Bante900P portable meter provides a quick manual calibration method for the conductivity calibration.
3.1 Record the cell constant value on the conductivity electrode (e.g. K=1.01).
3.2 Press and hold the
key for 3 seconds to enter setup menu, the display shows
“CELL”.
3.3 Press ENTER key to confirm, the meter enters the cell constant setting mode.