background image

BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER 

19 

 
 

Ion Calibration 

Bante900P portable meter allows 2 to 5 points calibration in the ion concentration mode. Available calibration points include 0.001, 0.01, 
0.1, 1, 10, 100, 1000 and 10000ppm, mg/L, mol/L or mmol/L. To obtain accurate measurement results, we recommend that you perform 
ion calibration and measurement at same temperature. 
1. Press MODE key until the meter shows 

  indicator. 

 

 

 

 

2. If necessary, select the concentration unit and ion valence (Monovalent or Divalent) in the setup menu. 
 

 

When default concentration unit has changed, the meter will show 

indicator “CAL” and flashing uninterruptedly.

 Press CAL key, the meter 

enters the calibration mode. 

 

 

3. Press CAL key, the meter shows 0.001ppm (or mg/L, mol/L, mmol/L). 

 

 

Summary of Contents for Bante900P

Page 1: ...Bante900P Portable Multiparameter Water Quality Meter Instruction Manual BANTE INSTRUMENTS LIMITED Technology to create the future...

Page 2: ...lectrolyte solution membrane cap and batteries Display Bante900P portable meter is equipped with a clear and bright LCD display that used to show measured values mode indicators and message indicators...

Page 3: ...Press the key to enter the calibration mode Press and hold the key to enter the setup menu MI Press the key to store current measured value Press key in setup mode to scroll up through menu Press key...

Page 4: ...e the battery cover from meter s backside 2 Insert the batteries into the battery compartment note polarity 3 Replace the battery cover into its original position Installation is completed When batter...

Page 5: ...Solutions Bante900P portable meter is packaged with three pH buffer packets required for calibration Open the pH7 00 buffer packet place the powder into a 250ml volumetric flask Pour distilled water...

Page 6: ...or tap water for at least 15 minutes DO NOT use distilled or deionized water it will shorten the life of sensor For the Ion Selective Electrodes Soak the electrode in the diluted standard solution e...

Page 7: ...ration Points Select the number of calibration points 1 point 2 points 3 points 4 points 5 points Resolution Sets the resolution for pH measurement 0 001pH 0 01pH 0 1pH Measurement Unit Sets the defau...

Page 8: ...e Coefficient Select the temperature coefficient of sample solution Range 0 0 10 0 2 10 Calibration Points Select the number of calibration points 1 point 2 points 3 points 4 points 5 points Pure Wate...

Page 9: ...grams per liter ppm Parts per million mmHg Pressure unit kPa Pressure unit GENERAL OPTIONS PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT Stable Criteria Sets the stability standard for measurement...

Page 10: ...nction allows user to restore the meter back to factory default parameters When this function is enabled all calibration values and selected system parameters will be lost or reset Enable Disable Sett...

Page 11: ...ess ENTER key to confirm the meter returns to measurement mode Setting is completed You can easily set each default parameter as per the above method Setting the Date and Time 1 Press and hold the key...

Page 12: ...easurement mode Setting is completed Temperature Compensation In order to get accurate measuring results you need to enable the manual or automatic temperature compensation before measurement or calib...

Page 13: ...to set the temperature value of sample 4 Press ENTER key to confirm the meter returns to measurement mode Setting is completed In the temperature setting mode press or key once the setting value will...

Page 14: ...ll not be accepted Make sure to calibrate the meter when attaching a new electrode or during first use DO NOT reuse calibration solution after calibration contaminants in solution will affect the cali...

Page 15: ...bration 2 3 Rinse the pH electrode with distilled water 2 4 Dip the electrode into the next standard buffer solution e g pH4 01 The meter automatically sense the current calibration solution and begin...

Page 16: ...tinue with second point calibration 3 5 Rinse the pH electrode with distilled water Dip the electrode into the next custom buffer solution 3 6 Press or key to increase or decrease displayed value pres...

Page 17: ...ows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format mm dd 4 Press key the meter shows OFS Zero point of...

Page 18: ...ss MODE key until the meter shows indicator 2 Rinse the ORP electrode with distilled water Dip the electrode into the calibration solution Stir the sensor gently 3 Press CAL key the meter shows curren...

Page 19: ...de 1 Press MR key in the ORP measurement mode the meter shows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date...

Page 20: ...te measurement results we recommend that you perform ion calibration and measurement at same temperature 1 Press MODE key until the meter shows indicator 2 If necessary select the concentration unit a...

Page 21: ...sured value to stabilize the display shows 0 1 CAL2 The meter prompts you to continue with second point calibration 7 Dip the Ion Selective Electrode into the corresponding calibration solution Stir t...

Page 22: ...in the ion measurement mode the meter shows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format mm dd 4 Pre...

Page 23: ...ccuracy we recommend that you perform 3 points calibration or select a standard value close to the sample value you are measuring The meter will automatically detect these conductivity standard soluti...

Page 24: ...r returns to measurement mode automatically Calibration is completed MULTI POINT CALIBRATION 2 1 Make sure that you have selected 2 to 5 points calibration in setup menu 2 2 Repeat steps 1 2 to 1 5 ab...

Page 25: ...ts you to continue with third point calibration 2 7 Repeat steps 2 3 to 2 5 above until the display shows END the meter returns to measurement mode automatically Calibration is completed MANUAL CALIBR...

Page 26: ...mode 3 6 Press or key to set each digit according to previous record press ENTER key to confirm each selection until the setting values stop flashing The meter returns to measurement mode automatical...

Page 27: ...measurement mode the meter shows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format mm dd 4 Press key the...

Page 28: ...saturation calibration in the air If you select 2 points calibration you need to use zero oxygen solution 100 SATURATION CALIBRATION 1 1 Press MODE key until the meter shows measurement unit 1 2 Make...

Page 29: ...ro oxygen solution stir the probe gently 2 5 Press ENTER key to confirm Calibration indicator begins flashing 2 6 Wait for the measured value to stabilize the display shows 8 25 CAL2 The meter prompts...

Page 30: ...linity and atmospheric pressure coefficient in the setup menu Refer to Setup Menu section 1 4 Press CAL key the meter shows current calibration point e g 8 25mg L 1 5 Dip the DO probe into the air sat...

Page 31: ...ro oxygen solution stir the probe gently 2 5 Press ENTER key to confirm Calibration indicator begins flashing 2 6 Wait for the measured value to stabilize the display shows 8 25 CAL2 The meter prompts...

Page 32: ...e temperature value 3 Press ENTER key to confirm Calibration is completed pH Measurement 1 Press MODE key until the meter shows indicator 2 Rinse the pH electrode thoroughly with distilled or de ionis...

Page 33: ...shows indicator 2 Rinse the Ion Selective Electrode thoroughly with distilled or de ionised water 3 Dip the electrode into the sample solution stir the sensor gently 4 Wait for the reading to stabiliz...

Page 34: ...ater brine and other liquids If you use meter to measure seawater or other water containing large amounts of salt please set salinity coefficient before use Some gas and steam such as chloride sulfur...

Page 35: ...immediately freeze currently displayed value Press the key again to release the measured reading you can continue to take measurement Storing and Recalling Data from Memory The meter allows up to 500...

Page 36: ...ER QUALITY METER 35 3 Press key the meter shows the date and time of the stored data set Format mm dd 4 Press key again the display shows the stored data 5 Press MEAS key the meter exits current mode...

Page 37: ...System icon on the start bar the DAS splash screen appears for few seconds and then the application starts RECEIVING THE STORED DATA Connect the USB cable to meter and computer Communication port COM...

Page 38: ...data If you need to view previous data click the Recall button the screen will show all data in the database DELETE THE DATA If you need to delete the certain data click on the data field then click...

Page 39: ...tton the screen will show a dialog box Selecting the time interval and recording time click the OK button the measured value will automatically transfer to data sheet CREATE THE EXCEL DATA FILE When t...

Page 40: ...f HCL Place the electrode in the solution for 10 minutes Rinse the sensor with distilled water REACTIVATING THE ELECTRODE If stored and cleaned properly the electrode should be ready for immediate use...

Page 41: ...r paper and install the probe Troubleshooting LCD DISPLAY CAUSE CORRECTIVE ACTION Electrode dried out Soak the pH electrode in 3M KCL solution for 10 minutes Soak the Ion Selective Electrode in the di...

Page 42: ...elective Electrode Accuracy 0 5 F S Monovalent 1 F S Divalent Resolution 0 001 0 01 0 1 1 Calibration Points 2 to 5 points Calibration Solutions 0 001 0 01 0 1 1 10 100 1000 10000ppm mol L mg L Conduc...

Page 43: ...s 1 or 2 points Temperature Compensation 0 50 C 32 122 F Pressure Correction 60 0 112 5kPa 450 850mmHg Salinity Correction 0 50g L Saturation of Oxygen Range 0 0 200 0 Accuracy 2 0 Resolution 0 1 Temp...

Page 44: ...cts Bread Dough Cement Cosmetics Dairy Products Education Fats Cream Field Use Fish Products Lab Flasks Low Ionic Meat Cheese Micro Samples Paint Photographic Soil Surface Test Tubes Tris Buffer Visco...

Page 45: ...phide Addendum 4 How to prepare the conductivity standard solutions Place AR potassium chloride reagent in a 50mL beaker and dry in an oven for 3 to 5 hours at 105 C then cool to room temperature in d...

Page 46: ...800 91 7 688 900 90 5 679 1000 89 4 671 1100 88 3 662 1200 87 2 654 1300 86 1 646 1400 85 0 638 1500 84 0 630 1600 82 9 622 1700 81 9 614 1800 80 9 607 1900 79 9 599 2000 78 9 592 Addendum 6 How to pr...

Page 47: ...hese efforts and we expect to have little or no such materials in our product in the coming years We welcome our customer suggestions on how to speed up these efforts Warranty The warranty period for...

Reviews: