BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER
20
4. Press
or
key to select the desired calibration point (e.g. 0.01ppm).
5. Dip the Ion Selective Electrode into corresponding calibration solution. Stir the sensor gently to create a homogenous solution.
6. Press ENTER key to confirm
, “Calibration”
indicator begins flashing. Wait for the measured value to stabilize, the display shows
“0.1/CAL2”
. The meter prompts you to continue with second point calibration.
7. Dip the Ion Selective Electrode into the corresponding calibration solution. Stir the sensor gently.
8
. Press ENTER key, “Calibration” indicator
begins flashing. Wait for the measured value to stabilize, the display shows
“1/CAL3”
. The meter
prompts you to continue with third point calibration.
9. Repeat steps 7 to 8
above until the display shows “END”
, the meter returns to measurement mode. Calibration is completed.